![Geokon 4500HT Скачать руководство пользователя страница 11](http://html.mh-extra.com/html/geokon/4500ht/4500ht_instruction-manual_2214561011.webp)
5
5.2 GK-405 Readout Box
The GK-405 Vibrating Wire Readout is made up of two components: The Readout Unit,
consisting of a Windows Mobile handheld PC running the GK-405 Vibrating Wire Readout
Application; and the GK-405 Remote Module, which is housed in a weatherproof enclosure and
connects to the vibrating wire gage to be measured. The two components communicate
wirelessly using Bluetooth
®
, a reliable digital communications protocol. The Readout Unit can
operate from the cradle of the Remote Module, or, if more convenient, can be removed and
operated up to 20 meters from the Remote Module.
Figure 4 - GK-405 Readout Box
5.2.1 Operating the GK-405
1)
Attach the GK-403-2 flying leads to the unit.
2)
Press the button labeled “POWER ON (BLUETOOTH)”.
3)
A blue light will begin blinking, signifying that the Remote Module is waiting to
connect to the handheld unit.
4)
Launch the GK-405 VWRA program handheld PC by tapping on “Start”, then
“Programs ”, and then the GK-405 VWRA icon.
5)
After a few seconds, the blue light on the Remote Module should stop flashing and
remain lit, indicating that the remote module has successfully paired with the
handheld PC.
6)
The red and black clips of the GK-403-2 flying leads are used for taking both the
pressure reading
and
the temperature reading. The white and green clips of the flying
leads are not used with the 4500HT. Make sure the shield drain wire is connected to
the blue clip on the flying leads for both readings.
7)
To take a pressure reading, connect the red and black clips from the flying leads to
the red and black leads of the 4500HT.
Содержание 4500HT
Страница 2: ......
Страница 4: ......
Страница 15: ...9 Figure 5 A Typical Model 4700 VW Temperature Sensor Calibration Report ...
Страница 16: ...10 Figure 6 Typical Calibration Report at 25 C ...
Страница 17: ...11 Figure 7 Typical Calibration Report at 100 C ...
Страница 18: ...12 Figure 8 Typical Calibration Report at 150 C ...
Страница 19: ...13 Figure 9 Typical Calibration Report at 200 C ...
Страница 20: ...14 Figure 10 Typical Calibration Report at 250 C ...