Pronto-250 User Manual 

Revision 2.2 

 

1.3. SPECIFICATIONS 

 
The following specifications are based on a one-year calibration cycle, an operating temperature of 15 to 
28ºC (59 to 82ºF) and a relative humidity not exceeding 80%. Monitors must be stored in an environment 
between 10

o

C to 60

o

C (50 to 140

o

F) and a relative humidity not exceeding 90%.

 

 

 

 

Pronto-250 

Interface 

Touch Screen 

Display 

Color LCD 28 X 35 mm (128 x 160 pixels) 

Measurement Controls 

Wavelength Selection and User Calibration 

Data Acquisition and Transfer 

On/Off Controls 

Maximum of 50 000 Measurements 

Screen Personalization 

4 screen orientations and 4 brightness levels 

Saved Settings 

Screen Orientation 

Screen Brightness 

Wavelength 

User calibration 

Effective Aperture Diameter 

19 mm Ø 

Spectral Range 

190 nm 

– 20 µm 

Calibrated Spectral Range 

0.248 

– 2.5 µm and 10.6 µm 

Power Noise Level 

10 mW 

Response Time 
(0 - 100 %) 

5 s 

Typical Sensitivity 

0.23 mV/W 

Calibration Uncertainty 

± 3 % 

Linearity with Power 

± 2 % 

Repeatability (Precision) 

± 0.5 % 

Maximum Power Resolution 

1 mW 

Minimum Power 

0.5 W 

Maximum Power 

250 W 

(A maximum exposition of 6 sec is allowed at 250 W)  

Pulsed Laser Damage Thresholds 

1.064 µm, 360 µs, 5 Hz 

1.064 µm, 7 ns, 10 Hz 

532 nm, 7 ns, 10 Hz 
266 nm, 7 ns, 10 Hz

 

Max. Energy Density 

9 J/cm

2

 

1.0 J/cm

2

                                                             

0.6 J /cm

2

 

0.3J /cm

2

 

Peak Power Density 

25 kW/cm

2

 

143 MW/cm

2

 

86 MW/cm

2

 

43 MW/cm

2

 

Minimum Frequency

 

15Hz

 

Maximum Number of Readings 
before probe must be cooled 
[for 25°C starting temp.] 

5 W: 50 

25 W: 10 

50 W: 5 

125 W: 2 
250 W: 1 

Maximum Exposure Time before 
probe must be cooled 
[for 25°C starting temp.] 

5 W: 120 sec 
25 W: 60 sec 
50 W: 30 sec 

125 W: 12 sec 

250 W: 6 sec 

Maximum Device Temperature 

65°C 

Operating Conditions 

Ambient temperature : 15-28°C 

Maximum Relative humidity : 80% 

Содержание Pronto-250

Страница 1: ...Pronto 250 User Manual Revision 3 0 i...

Страница 2: ...er Support Gentec Electro Optics Inc 445 St Jean Baptiste Suite 160 Qu bec QC Canada G2E 5N7 Tel 418 651 8003 Fax 418 651 1174 E mail service gentec eo com Website gentec eo com Lifetime Warranty Gent...

Страница 3: ...und to comply with the limits for a Class A digital device pursuant to part 15 of the FCC Rules These limits are designed to provide reasonable protection against harmful interference when the equipme...

Страница 4: ...ing Transferring and Deleting Data 11 3 4 1 Acquiring the Data 11 3 4 2 Transferring the Data 11 3 4 3 Deleting the data 11 4 Safety Instructions 12 4 1 General 12 4 2 Damage to the Optical Absorber M...

Страница 5: ...BEAM IMMEDIATELY TO AVOID DAMAGING THE SENSOR Be careful not to exceed the maximum levels and densities stated in the specifications The handle on the Pronto 250 is for carrying purposes only The user...

Страница 6: ...l Range 190 nm 20 m Calibrated Spectral Range 0 248 2 5 m and 10 6 m Power Noise Level 10 mW Response Time 0 100 5 s Typical Sensitivity 0 23 mV W Calibration Uncertainty 3 Linearity with Power 2 Repe...

Страница 7: ...sidered as the safe minimum diameter Figure 1 Minimum Beam Size for Pronto 250 Absorber at 1064 nm Max Power Density 1 064 m CW 10 6 m CW 45 kW cm2 14 kW cm2 Dimensions 59 0W x 181 4L x 17 3D mm Weigh...

Страница 8: ...r a safe use during the measurements Touch Screen Display and Controls The touch screen interface controls the device On Off Settings Button This button is used to turn the device On press when the de...

Страница 9: ...cquisition Press Starts Data Acquisition Press Stops Data Acquisition If gray Function is inactive UP DOWN Press Increments the selected units Press Decrements the selected units About Press Opens the...

Страница 10: ...Battery indicators Ready Dots appear in sequence while the device waits for a trigger signal In this state the Zero is continuously updated to ensure the accuracy of the measurement Countdown A live C...

Страница 11: ...ration Displays the current Correction Factor applied to the measurements and the corresponding corrected power to help the user adjust the value Displays Up Down controls to adjust the correction fac...

Страница 12: ...blue and use the Down button to reduce it to 0 Select the 2nd number Make sure it turns blue and use the Up button to increase it to 5 Select the 3rd number Make sure it turns blue and use the Down b...

Страница 13: ...it turns blue like this and then increment it twice with the Up button Initial Setting Select the Digit Change the Digit using the Up Button The last measured value 145 3 W is displayed between the Up...

Страница 14: ...to its initial state displaying the last measured value 3 Place the device in the laser beam path Be careful to conform to all the recommended specifications for beam size placement and laser power 4...

Страница 15: ...to your computer with a USB cable and use the ProntoDataTransfer software You can download our latest version of the software in the Downloads section of our website https gentec eo com downloads The...

Страница 16: ...to 80 of the sensor s aperture For the Pronto 250 this corresponds to a diameter of 16 17 mm In any case the beam s incident area should not be less than 10 of the detector s area Please contact Gente...

Страница 17: ...ing holes with an 8 32 thread see Figure 2 Figure 3 Recommended setup when making a measurement install the device on an optical stand before placing it in the laser beam path Gentec EO offers a compa...

Страница 18: ...device s firmware are created it is in your best interest to update your PRONTO 250 The latest device firmware can be downloaded from the Gentec EO website at https gentec eo com downloads Go to the D...

Страница 19: ...is declared Emissions Product Standard Test Standard Description EN 61326 1_Ed2 2013 IEC 61326 1_Ed2 2012 CISPR 11 A1 2010 Class A Radiated Emissions EN 61326 1_Ed2 2013 IEC 61326 1_Ed2 2012 Radiated...

Страница 20: ...ter when the detector reaches the end of its life Removing the insulation or troubling the inside of the monitor will void the detector warranty The complete Detector contains 1 Detector 1 cable 1 cal...

Страница 21: ...website 2 Download the software file from our website 3 Double click the EXE file to start the installer 4 Choose the folder for the installation and click Next until the installation is complete The...

Страница 22: ...a Download File button 7 Click the Download File button and select a known folder on your computer where you will save the data Don t forget to also enter a file name The file format is TXT 8 You can...

Страница 23: ......

Отзывы: