Example
Command: *GZO
Answer:
Zero Value = 1619.352000 W<CR><LF>
*MUL - Set User Multiplier
This command is used to set the value of the user-specified multiplier. Note, the multiplier cannot be set at
0.
Command
Parameters
Answer
MUL
8-character numerical value
Multiplication factor
Default
: 1
Example
The following example sets multiplier = 33
Command: *MUL00000033
Or
*MUL3.3000e1
Answer:
Multiplication factor = 33.000000<CR><LF>
*GUM - Get User Multiplier
This command returns the multiplier value.
Command
Parameters
Answer
GUM
None
Current multiplier value
Example
Command: *GUM
Answer:
Multiplication factor = 33.000000<CR><LF>
*OFF - Set User Offset
This command is used to set the value of the user-specified offset.
Command
Parameters
Answer
OFF
8-character numerical value
None
Default:
0
Example