HP-BLU User Manual
Revision 1.0
14
Example
The following example sets the wavelength to 25 microns (25000 nm).
Command: *PWM025.0
Answer: Wavelength = 25000.000000 nm,
factor = 1.000000<CR><LF>
*GWL - Get Wavelength
This command returns the wavelength setting in nm.
Command
Parameters
Answer
GWL
None
Returns the wavelength in nm
Example
Command: *GWL
Answer: 1064.000000 nm<CR><LF>
4.5.3. Control
*ANE - Set Anticipation
This command is used to enable the anticipation algorithm. Please see section 1.3.9 for more details on
the anticipation algorithm and how it affects the power measurement.
Command
Parameters
Answer
ANE
None
None
Default:
On
Example
The following example sets the anticipation On.
Command: *ANE
Answer:
*AND
– Clear Anticipation
This command is used to disable the anticipation algorithm. Please see section 1.3.9 for more details on
the anticipation algorithm and how it affects the power measurement.
Command
Parameters
Answer
AND
None
None
Default:
On