
21
WARNING
Hazardous voltages exist at the outputs and the load connections. To protect personnel against
accidental contact with hazardous voltages, ensure that the load and its connections have no
accessible live parts. Ensure that the load wiring insulation rating is greater than or equal to the
maximum output voltage of the power supply.
Fig. 3-6: Load wires connection, 8V to 100V models.
Fig. 3-7: Bus-bars shield mounting,
8V to 40V models
Fig. 3-8: Bus-bars shield mounting,
60V to 100V models
150V to 600V Models
M10x25 screw (2 places)
Hex Nut (2 places)
Flat washer (2 places)
Spring washer (2 places)
Wire terminal lug (2 places)
Flat washer
(2 places)
Screws tightening torque: 290-310 Lb-inch.
Output Load Wires No Conductor Pretreatment: Phoenix Contact clamping parts are designed
so that all kinds of copper conductors can be clamped without pretreatment.
It is forbidden to solder the conductors. The solder tin yields and fractures under high pressure. The
result is increased contact resistance and an excessive temperature rise. In addition, corrosion
caused by pickling or fluxes has been observed on soldered conductor ends. Notch fractures at the
transition point from the rigid to the flexible conductor area are also possible.
CAUTION
21
Содержание GEN10-330
Страница 3: ...This page intentionally left blank...
Страница 4: ......
Страница 9: ...This page intentionally left blank...
Страница 40: ...31...
Страница 42: ...33...
Страница 59: ...50...
Страница 77: ...NOTES 68...
Страница 78: ...NOTES 69...
Страница 79: ...NOTES 70...
Страница 80: ...NOTES 71...
Страница 81: ...NOTES 72...