8
Control solution is used to check the performance of the kit.
The meter kit should be check:
• When the meter and strip do not work properly.
• When the test result is unusual or inconsistent.
Control solution range is shown as follow:
Please check your strip vial label for exact range.
Perform a Control Test:
Insert an Unused Strip
→
Make Sure the Code Number is Correct (Total Cholesterol/
Uric Acid)
→
Press “M” Button (3 secs)
→
Enter Control Solution Mode (Shown CL
on Screen)
→
Shake the Control Solution Well
→
Discard First Three Drops
→
Put
One Drop onto a Clean Surface
→
Touch the Control Solution with Strip Sample
Inlet
→
Test Result Display after Count Down
→
Compare the Result with the Range
Listed on Test Strip Vial
Note:
• Do not reuse the test strip.
• Marked open date on new open control solution.
• Do not use the meter if the control test is out of range.
• If the control test keeps result wrongly, please contact an authorized distributor.
• Control solutions are not included. Please contact an authorized distributor for purchasing.
Chapter 5
Control Solution Test
98-164 mg/dL
5.4-9.1 mmol/L
EA13120901
Control Range LV1
2017-01
Test Strip
7142
Operatio n
Storage &
Transportation
Storage &
Transportation
90%
10%
98-164 mg/dL
5.4-9.1 mmol/L
G1303190
2017/01
EA13120901
2017-01