DATABASE IMPORT / EXPORT (Setup Mode P8)
Database Import / Export
GA-Series balances allow you to Import and Export to a Flash Drive via the USB 1 Port on the back of
the balance. Scroll to Setup Level P8 or insert a flash drive into the USB 1 port, IEI IMPORT / EXPORT
appears on the display. Press the Enter Key, Export (IE1) appears on the display. Press the Up/Down
Arrow key to access the Import (IE2) data option. Press Enter Key followed by UP/Down Arrow
Keys
to view the following:
ALL DATABASES (IE1.1 – Export or IE2.1 - Import)
USERS (IE1.2 – Export or IE2.2 - Import)
PRODUCTS (IE1.3 – Export or IE2.3 - Import)
TARE (IE1.4 – Export or IE2.4 - Import)
WEIGH DATA (IE1.5 – Export or IE2.5 - Import)
ALIBI (IE1.6 – Export or IE2.6 - Import)
PARAMETERS (IE1.7 – Export or IE2.7 - Import)
PERIPHERAL DEVICES
Peripheral devices connected to the balance via the USB or RS232 Ports must be power in such a manner that
prevents the occurrence of different potentials in the cables. Communication parameters (Baud Rate & Parity)
of the balance and external device must correspond to one another.
Print/Send Commands:
-
Manual: On Pressing the Print/Enter Key
-
Automatic: Upon Reaching a Stable Weighing Result
-
Continuously: Continuous
Stability Criteria:
-
Stable: Upon Reaching a Stable Weighing Result
-
Unstable: Immediate Transmission of Weigh Value
DATA FORMAT
Data format sent from the balance to a peripheral device, 18 Characters:
-
Stability Marker:
[Space] When Stable
[?] When Unstable
[
∧
] Sample Exceeds Maximum
[
∨
] Sample Exceeds Maximum Weight –
-
Character:
[Space] – For Positive Value or [-] For Negative Value
-
Mass:
9 Characters, Aligned to the Right
-
Unit:
3 Characters, Aligned to the Left