and to prolong useful life of the probe tip.
Periodic Re-Calibration
– Always use plumb 14K
YELLOW gold to calibrate the AGT1 and make sure
you file the area of the calibration piece prior to
performing this function and touch the pen probe tip
to the filed area. Testing gold and gold-filled or base
metals will cause slight chemical changes in the
pen probe. To adjust for this chemical change, the
calibration process is needed. To ensure accurate
readings, recalibrate the unit from time to time and
wipe the pen probe tip to clean it with a dry paper
towel each time prior to recalibrating.
Effects of Alloys
– When alloyed with gold, COPPER
and NICKEL will drive readings downwards while
PALLADIUM, PLATINUM and SILVER will drive
readings upwards.