5. It is recommended that you keep a set of 10K,
14K, 18K, 22K and 24K reference gold samples
nearby to compare against the gold you are testing
if a questionable test occurs.
6. Always file the piece of metal being tested prior to
performing the test and then touch the pen probe
tip to the filed area. Be aware that in rare cases
gold plated, gold-filled and stainless steel jewelry
may react as karat gold or platinum unless filed
prior to testing.
7. The AGT1 is not user serviceable other than
battery replacement and probe cleaning or probe
replacement. If service is required, please contact
your supplier or the factory. Any attempt to repair
the tester will void the warranty.
TROUBLESHOOTING GUIDELINES:
TESTER START-UP
Warm-Up Time
– For best results, allow a 2 minute
warm-up time PRIOR to performing the initial
calibration of the tester. If using the optional AC
adaptor, after the AGT1 has warmed-up and been
calibrated, you may keep the unit turned on with no
fear of damaging the electronics and this will allow
you to avoid the need for ongoing warm up time
delays.
Excess Solution
– Fresh pen probes may be slightly
overfilled when first used and excess solution could
come out during the testing process. Simply dab the
pen probe tip on a dry paper towel a few times to
remedy this.
Test Pen Probe
– Prior to calibrating, touch and hold