Chapter 3 - System Setup
3-53
P R E L I M I N A R Y
NOTE:
This check uses both Ping and DICOM Ping. Ping is used to verify the TCP/IP connection to the server.
DICOM Ping is used to verify that the DICOM application is answering.
1) Select
DicomQR
from the Selected devices list.
2) Select Check.
3-8-6-3-1
If Retrieve is failing …
If Retrieve is failing, but Query is functioning, verify if the scanner has been registered as a receiver on
the server.
Содержание Venue Go
Страница 13: ...DIRECTION 5813707 100 REVISION 2 VENUE SERVICE MANUAL Table of Contents 13 P R E L I M I N A R Y...
Страница 14: ...DIRECTION 5813707 100 REVISION 2 VENUE SERVICE MANUAL 14 Table of Contents P R E L I M I N A R Y...
Страница 131: ...DIRECTION 5813707 100 REVISION 2 VENUE GO SERVICE MANUAL Chapter 3 System Setup 3 57 P R E L I M I N A R Y...
Страница 132: ...DIRECTION 5813707 100 REVISION 2 VENUE GO SERVICE MANUAL 3 58 Chapter 3 System Setup P R E L I M I N A R Y...
Страница 133: ...DIRECTION 5813707 100 REVISION 2 VENUE GO SERVICE MANUAL Chapter 3 System Setup 3 59 P R E L I M I N A R Y...
Страница 134: ...DIRECTION 5813707 100 REVISION 2 VENUE GO SERVICE MANUAL 3 60 Chapter 3 System Setup P R E L I M I N A R Y...
Страница 135: ...DIRECTION 5813707 100 REVISION 2 VENUE GO SERVICE MANUAL Chapter 3 System Setup 3 61 P R E L I M I N A R Y...
Страница 136: ...DIRECTION 5813707 100 REVISION 2 VENUE GO SERVICE MANUAL 3 62 Chapter 3 System Setup P R E L I M I N A R Y...
Страница 137: ...DIRECTION 5813707 100 REVISION 2 VENUE GO SERVICE MANUAL Chapter 3 System Setup 3 63 P R E L I M I N A R Y...
Страница 138: ...DIRECTION 5813707 100 REVISION 2 VENUE GO SERVICE MANUAL 3 64 Chapter 3 System Setup P R E L I M I N A R Y...
Страница 323: ......