Programming Site Data
2-19
March 2004
Entering Error Limits
The Error Limits option enables you to set limits for an incoming
signal. When the signal falls outside the programmed limits, an error
indication appears on the display screen. To set error limits, follow
the steps below.
Note:
If you are in the PROG menu, proceed directly to Step 4. If
you scroll to a different option, press [ENT] twice to select
that option (once to enter and again to confirm the selection).
1.
Press [ESC]. The UTX878 enters the
User Program
.
2.
Press the [
X
] key until
PROG
is bracketed in the upper left corner
and press [ENT].
3.
Use the [
S
] and [
T
] keys to scroll to the desired
Channel
and
press [ENT].
4.
Scroll to the
Error Limits
option and press [ENT].
5.
The first prompt asks for the
Minimum Signal
limit for the
transducer signal received by the UTX878. The E1:LOW
SIGNAL error message appears if the signal strength falls below
the limit programmed here. Press [ENT]. Use the arrow keys to
change the value and press [ENT].
6.
Repeat the steps above for the
Maximum Signal
limit.
7.
The next prompt calls for the
Minimum Velocity
limit. Press
[ENT]. Use the arrow keys to change the value and press [ENT].
8.
Repeat the steps above to change the
Maximum Velocity
limit.
(The E3: VELOCITY RANGE error message appears if the
velocity falls outside the minimum and maximum limits.)
9.
The meter now asks for the
Min Amplitude,
the lower limit for the
amplitude discriminator. The discriminator measures the size of
the transducer signal sent from the UTX878. If the signal falls
outside these limits, the E5: AMPLITUDE error message appears.
Press [ENT]. Use the arrow keys to change the value and press
[ENT].
10.
Repeat these steps for the
Max Amplitude
, the upper limit for the
discriminator.
11.
The next prompt asks for the acceptable limits for the sound
speed, based on conditions in your particular system. The E2:
SOUND SPEED error message appears if the fluid sound speed
exceeds that entered in the Fluid submenu by more than this
percentage. Press [ENT]. Use the arrow keys to change the
percentage and press [ENT].
Содержание UTX878 Ultrasonic
Страница 1: ...March 2004 Process Control Instruments UTX878 Ultrasonic Flow Transmitter User s Manual 910 253A ...
Страница 2: ...iii March 2004 The UTX878 Flow Transmitter and Transducers ...
Страница 8: ...Chapter 1 ...
Страница 20: ...March 2004 Installation 1 11 Figure 1 5 UTX878 Outline and Installation Dwg 712 1196 rev 2 ...
Страница 21: ...March 2004 Installation 1 12 Figure 1 6 Clamping Fixture Outline and Installation Dwg 570 076 rev 1 ...
Страница 22: ...March 2004 Installation 1 13 Figure 1 7 UTX878 Wiring Diagram Dwg 702 528 rev 2 ...
Страница 23: ...Chapter 2 ...
Страница 45: ...Chapter 3 ...
Страница 52: ...Chapter 4 ...
Страница 60: ...Chapter 5 ...
Страница 65: ...Chapter 6 ...
Страница 75: ...Chapter 7 ...
Страница 80: ...Appendix A ...
Страница 85: ...Appendix B ...
Страница 86: ...CE Mark Compliance Introduction B 1 EMC Compliance B 1 ...
Страница 88: ...Appendix C ...
Страница 89: ...Data Records Site Data C 1 ...
Страница 93: ...Appendix D ...
Страница 94: ...Service Record Introduction D 1 Data Entry D 1 Diagnostic Parameters D 3 ...
Страница 96: ...March 2004 D 2 Service Record Table D 1 Service Record cont Date Description of Service Performed By ...