9-8
Transport Pro
Revision A
2012659-042
Checkout: Electrical Safety Tests
powered from 100-240 V/50-60 Hz
11. Set the power switch of the device under test to OFF.
Patient (Source) Leakage Current
This procedure only applies to Class I (grounded/earthed) equipment,
and measures the leakage current from the ECG/RESP connector of the
device to ground.
1. Configure the leakage tester like the circuit shown below with GND
switch OPEN and polarity switch NORM.
2. Connect an ECG test body to the ECG/RESP connector of the device
under test.
3. Set the power switch of the device to ON.
4. Read the leakage current indicated on the DMM.
5. Change the leakage tester polarity switch to the RVS position.
6. Read the leakage current indicated on the DMM.
NOTE
If either reading is greater than 50 µA (0.05 volts on the DMM), the
device fails this test. Contact GE Technical Support.
7. Change the GND switch to the Closed position.
8. Read the leakage current indicated on the DMM.
9. Change the leakage current switch to the RVS position.
10. Read the leakage current indicated on the DMM.
NOTE
If either reading is greater than 10 µA (0.01 volts on the DMM), the
device fails this test. Contact GE Technical Support.
11. Set the power switch of the device to OFF.
NOTE
The AAMI and EN single fault condition (ground open) is 50 µA,
Power Cord
Leakage Tester
NORM
RVS
ECG Test Body
HIGH
LOW
GND
1K
0.15µF
10
Power Cord
Device
Under
Test
DMM set to measure AC voltage
GND
DMM
Closed
006A
Содержание Transport Pro
Страница 1: ...Transport Pro Patient Monitor Service Manual Software Version 2 2024579 018 Revision A ...
Страница 9: ...Revision A Transport Pro 1 1 2012659 042 1 Introduction ...
Страница 12: ...1 4 Transport Pro Revision A 2012659 042 Introduction Safety Message Definitions ...
Страница 13: ...Revision A Transport Pro 2 1 2012659 042 2 Equipment Overview ...
Страница 28: ...2 16 Transport Pro Revision A 2012659 042 Equipment Overview Theory of Operation ...
Страница 29: ...Revision A Transport Pro 3 1 2012659 042 3 Installation ...
Страница 33: ...Revision A Transport Pro 4 1 2012659 042 4 Configuration ...
Страница 38: ...4 6 Transport Pro Revision A 2012659 042 Configuration Monitor Settings ...
Страница 39: ...Revision A Transport Pro 5 1 2012659 042 5 Maintenance ...
Страница 47: ...Revision A Transport Pro 6 1 2012659 042 6 Troubleshooting ...
Страница 57: ...Revision A Transport Pro 7 1 2012659 042 7 Ordering Parts ...
Страница 59: ...Revision A Transport Pro 7 3 2012659 042 Ordering Parts Exploded View Exploded View 622A ...
Страница 62: ...7 6 Transport Pro Revision A 2012659 042 Ordering Parts Hardware ...
Страница 63: ...Revision A Transport Pro 8 1 2012659 042 8 Disassembly ...
Страница 83: ...Revision A Transport Pro 9 1 2012659 042 9 Checkout ...
Страница 97: ...Revision A Transport Pro A 1 2012659 042 A Technical Specifications ...
Страница 100: ...A 4 Transport Pro Revision A 2012659 042 Technical Specifications External Power Supply ...
Страница 101: ...Revision A Transport Pro B 1 2012659 042 B Electromagnetic Compatibility ...
Страница 109: ......