Electromagnetic Compatibility (EMC)
32
TONOPORT VI
2001589-312 Revision B
Guidance and Manufacturer’s Declaration—Electromagnetic Immunity
TONOPORT VI is intended for use in the electromagnetic environment specified below. It is the responsibility of the
customer or user to ensure that TONOPORT VI is used in such an environment.
Immunity Test
EN/IEC 60601 Test Level
Compliance
Level
Electromagnetic environment—
Guidance
Electrostatic discharge
(ESD) to EN 61000-4-2/
IEC 61000-4-2
±8.0 kV contact
±2.0 kV air
±4.0 kV air
±8.0 kV air
±15.0 kV air
±8.0 kV
±2.0 kV
±4.0 kV
±8.0 kV
±15.0 kV
Floors should be wood, concrete or
ceramic tile. If floors are covered with
synthetic material, the relative humidity
should be at least 30%.
Electrical fast transient/
burst to EN 61000-4-4/
IEC 61000-4-4
±2.0 kV for power supply
lines
±1.0 kV for input/output
lines
not applicable
not applicable
Mains power should be that of a typical
commercial or hospital environment.
Surge to EN 61000-4-5/
IEC 61000-4-5
±0.5 kV differential mode
±1.0 kV differential mode
±0.5 kV common mode
±1.0 kV common mode
±2.0 kV common mode
not applicable
not applicable
Mains power should be that of a typical
commercial or hospital environment.
Voltage dips, short inter-
ruptions and voltage varia-
tions on power supply
input lines to EN 61000-4-
11/IEC61000-4-11
0 % power supply
for 10 ms (0.5 cycles)
0 % power supply
for 20 ms (1.0 cycle)
70 % power supply
for 500 ms (25 cycles)
0 % power supply
for 5000 ms (250 cycles)
not applicable
not applicable
not applicable
not applicable
Mains power should be that of a typical
commercial or hospital environment. If
the user of TONOPORT VI requires con-
tinued operation during power mains
interruptions, it is recommended that
TONOPORT VI be powered from an
uninterruptible power supply or a battery.
Power frequency (50/
60 Hz) magnetic field to
EN 61000-4-8/IEC 61000-
4-8
30.0 A/m
30.0 A/m
Power frequency magnetic fields should
be at levels characteristic of a typical
location in a typical commercial or
hospital environment.
Содержание TONOPORT VI
Страница 38: ...Index 38 TONOPORT VI 2001589 312 Revision B U USB driver installation 26 W Warning 5 Weight 29 ...
Страница 39: ......