Controls and Indicators
2001589-312 Revision B
TONOPORT VI
11
2
Controls and Indicators
Fig. 2-1 Controls and indicators of TONOPORT VI
Functions of Button
1
Button
:
push to display the most recent pa-
rameter readings. The display will show:
- systolic value "S" (unit mmHg or kPa shown on the
display)
- diastolic value
"
D
"
(unit mmHg or kPa shown on
the display)
- pulse rate "HR" (unit min
-1
)
The same button is used
- to toggle between the day phase and the night
phase (section "Toggling Between Day and Night
Phase") and
- to program the BP monitor (chapter 3 "Setup")
2
Connection for blood pressure cuff
3
Calibration mark
4
Liquid crystal display (LCD)
5
Button
:
push to start and stop a measure-
ment and to confirm entries
6
(Rechargeable) batteries
7
Lid covering battery compartment
8
Nameplate
9
Port for connection to PC (RS232)
10
Port for connection to PC (USB)
1
2
3
4
5
6
7
8
9
10
Button
Mes-
sage on
display
Function
Push once
H 1
clear memory
Push twice
H 2
set date and time
Push 3 times
H 3
select measurement protocol
Push 4 times
H 4
activate calibration mode
Push 5 times
H 5
display firmware version
Push 6 times
H 6
select energy source
Push 7 times
H 7
enable/disable audio signal
Push 8 times
H 8
toggle pressure unit between
mmHg and kPa
Push 9 times
H 9
select measurement method:
deflation measurement
method or inflation
measurement method
INFO
INFO
INFO
START
STOP