Chapter 1. Getting Started
Model PM880 Portable Hygrometer User’s Manual
13
1.4.2b Entering Moisture Calibration Data (cont.)
Figure 5: Hygro Window in Calibration
7.
Press [
] to move to the next text box, and press
[ENTER]
. Use the
numeric keys to enter the corresponding dewpoint reading and press
[ENTER]
.
8.
Use the arrow keys to scroll to additional data points and repeat steps 5,
6 and 7 until you have entered the values for each data point.
Note:
To insert an additional data point, press
[F1]
(
Insert
).
To delete a data point, press
[F2]
(
Delete
).
9.
When you have completed entering values, press
[F3]
(
EXIT
).
Next, do one of the following:
•
If the probe has a pressure transducer, proceed to step 3 in
Entering
Pressure Calibration Data
on page 14.
•
If the probe does not have a pressure transducer, press
[F3]
(
OK
).
Proceed to
Displaying Measurements
on page 15.
Содержание PM880
Страница 2: ......
Страница 4: ...ii no content intended for this page ...
Страница 14: ...Contents xii Model PM880 Portable Hygrometer User s Manual no content intended for this page ...
Страница 34: ...Model PM880 Portable Hygrometer User s Manual 20 Chapter 1 Getting Started no content intended for this page ...
Страница 108: ...Chapter 5 Logging Data 94 Model PM880 Portable Hygrometer User s Manual no content intended for this page ...
Страница 196: ...Appendix B Menu Maps 182 Model PM880 Portable Hygrometer User s Manual no content intended for this page ...
Страница 202: ...Model PM880 Portable Hygrometer User s Manual 188 Appendix B Menu Maps no content intended for this page ...
Страница 212: ...Index 198 Model PM880 Portable Hygrometer User s Manual no content intended for this page ...
Страница 217: ......