Chapter 9. Features and Specifications
172
Model PM880 Portable Hygrometer User’s Manual
9.7 Moisture Measurement
Compatibility
Compatible with all GE aluminum-oxide moisture probes:
M Series, TF Series and Moisture Image Series. Each probe type requires a
different type of I/O cable.
Calibration
GE moisture sensors are computer calibrated to NIST-traceable moisture
concentrations.
9.8 Dew/Frost Point Temperature
Overall Range Capability
•
–110° to 60°C (–166 to 140°F)
Available Range Options
•
Standard: –80° to 20°C (–112° to 68°F) with data to –110°C (–166°F)
•
Ultra-Low: –110° to –50°C (–166° to –58°F)
•
High Range Data: –80° to 60°C (–112° to 140°F)
Calibrated Accuracy at 25°C (77°F)
•
±2°C (3.6°F) from –65° to 10°C (–85° to 50°F)
•
±3°C (5.4°F) from –80° to –66°C (–112° to –87°F)
Repeatability
•
±0.5°C (0.9°F) from –65° to 10°C (–85° to 50°F)
•
±1.0°C (1.8°F) from –80° to –66°C (–112° to –87°F)
Operating Pressure
5 µ of Hg to 5,000 psig (345 bar) limited by pressure sensor —
(see pressure measurement specifications)
Содержание PM880
Страница 2: ......
Страница 4: ...ii no content intended for this page ...
Страница 14: ...Contents xii Model PM880 Portable Hygrometer User s Manual no content intended for this page ...
Страница 34: ...Model PM880 Portable Hygrometer User s Manual 20 Chapter 1 Getting Started no content intended for this page ...
Страница 108: ...Chapter 5 Logging Data 94 Model PM880 Portable Hygrometer User s Manual no content intended for this page ...
Страница 196: ...Appendix B Menu Maps 182 Model PM880 Portable Hygrometer User s Manual no content intended for this page ...
Страница 202: ...Model PM880 Portable Hygrometer User s Manual 188 Appendix B Menu Maps no content intended for this page ...
Страница 212: ...Index 198 Model PM880 Portable Hygrometer User s Manual no content intended for this page ...
Страница 217: ......