Chapter 6: Illustrated Parts
Test Points
6-36
Service Manual
6.12 Test Points
The DC voltage generated on Control and Power PCAs shall meet the following requirements:
•
The following are derived from the 3.3V input voltage: 1.8V ± 3%, 2.5V ± 2%, 3.3V+0.2V/-0V
•
The following are derived from the 12V input voltage: ISO5V ± 5%, 5.0V ± 4%, 3.3VSTBY ± 2%, 12V ±
5%, 5V -ni ± 5%
6.12.1 Control Board
TP 1, 2
TP Value
1 +3.3V
(analog)
2 Ground
(analog)
3 +2.5V
(digital)
4 Ground
(digital)
5 +1.8V
(digital)
6 +5V
(isolated)
10 Ground
(isolated)
7 +12V
(digital)
8 Ground
(digital)
9 +3.3V
(digital)
11 Ground
(digital)
12 +5V
(digital)
13 +3.3V
(standby
TP 3, 4, 5
TP 7, 8, 9
TP 6, 10
TP 11, 12
TP 13