6 Reference and Specification
K0443 Revision A
6 - 28
[EN] English
Program
To start a test program do the following:
1.
Select the test programs listed on the screen.
2.
The step function moves down the list of available test programs.
3.
To start a test program, press run.
4.
When the program starts, a stop legend replaces the run legend.
5.
Press the stop key at any time to stop the test program.
To write a test program do the following:
1.
Press Set-up.
2.
Select New.
Note:
A new test always contains a START and END command (first and last line instructions),
these lines cannot be deleted.
To edit an existing program do the following:
1.
Press Edit.
To enter a name for the program do the following:
1.
Use g the text editor keys to select letter range to select the letter within the range.
2.
Use the right arrow key to move to the next character position.
3.
Press enter on completion.
The display shows the line instruction with insert and delete selections. Pressing insert
changes the display to a list of the available programming commands.
Test programs can be saved to or uploaded from a USB stick.
Test programs can also be edited or written on a PC.