PACE Pressure Controller User Manual
[EN] English
5 - 1
K0443 Revision A
5
Testing and Fault Finding
5.1
Introduction
This section details the standard serviceability test.
(Ref: Fault Diagnosis,
Table 5.1)
for
possible faults, and the response.
The PACE instrument contains a self-test and diagnosis system that continuously monitors
the performance of the unit. At power-up, the system performs a self-test.
5.2
Standard Serviceability Test
The following procedure shows if the unit is serviceable and checks functions and facilities of
the PACE instrument.
Procedure
1.
Connect the instrument
(Ref: Installation, Section 2)
Connect a pressure measurement
device to the output port.
2.
After power-up, select measure set-up.
a.
Select the required units of pressure measurement from the measure set-up
menu.
b.
Press the status area to enter control set-up.
c.
Select the effort meter.
d.
Select the required control slew rate and vent rate.
e.
Press set-point and, using the numeric keys, set a value within the pressure
range of the instrument.
f.
Check the screen display shows the following:
•
Selected units of pressure measurement
•
Selected type of slew rate
•
Set-point
g.
Press the
key to start
3.
The screen display changes as follows:
a.
The measured pressure digits change from black to blue and indicates the
pressure value changing towards the set-point.
b.
If enabled, the effort meter indicates the work effort of the controller.
Control
Содержание PACE5000
Страница 9: ...K0443 Revision A viii EN English 6 14 Vacuum System Parts 6 43 ...
Страница 49: ...4 Maintenance K0443 Revision A 4 6 EN English Intentionally Blank ...
Страница 53: ...5 Testing and Fault Finding K0443 Revision A 5 4 EN English Intentionally Blank ...
Страница 97: ...6 Reference and Specification K0443 Revision A 6 44 EN English Intentionally Blank ...