PACE Pressure Controller User Manual
[EN] English
2 - 5
K0443 Revision A
Supply equipment
Pneumatic supplies should have isolation valves and, where necessary, conditioning
equipment.
The positive pressure supply should be regulated to between 110% of the full-scale pressure
range and MWP stated on the control module.
To protect the instrument from over-pressure a suitable protection device (such as a relief
valve or bursting disc) must be fitted to prevent over pressurization.
On instruments without a negative supply, the positive pressure discharges from the system
to atmosphere through the negative supply port. Pipe the negative port to a safe discharge
area or fit a diffuser to the negative port.
During system pressure vent operations, the pressure discharges from the system to
atmosphere through the negative and vent ports. Pipe both ports to a safe discharge area or
fit a diffuser to the negative port.
Pneumatic Connection Examples
(Ref: Figures 2-3, 2-4 and 2-5)
The examples that follow show a single channel connection detail, using supply equipment
described above.
Figure 2-3, Pneumatic Connections without vacuum supply
1) Pressure source
2) Conditioner
3) Filter
4) Regulate to between 110% full-scale and MWP
5) Diffuser*
6) Unit under test
7) Optional reservoir
†
8) Protection device
9) Optional differential connection
14) Manual external vent valves
a) Atmosphere
OUTPUT
1
2
3
4
5
6
7
+
5
9
a
-
SUPPLY
VENT REF
8
14
14
Содержание PACE5000
Страница 9: ...K0443 Revision A viii EN English 6 14 Vacuum System Parts 6 43 ...
Страница 49: ...4 Maintenance K0443 Revision A 4 6 EN English Intentionally Blank ...
Страница 53: ...5 Testing and Fault Finding K0443 Revision A 5 4 EN English Intentionally Blank ...
Страница 97: ...6 Reference and Specification K0443 Revision A 6 44 EN English Intentionally Blank ...