PACE Pressure Controller User Manual
[EN] English
5 - 3
K0443 Revision A
5.3
Fault Finding
Check the faults and responses
(Ref: Fault Diagnosis, Table 5.1)
before contacting
www.gemeasurement.com
or a GE recommended Service Agent
Table 5.1 - Fault Diagnosis
5.4
Approved Service Agents
For the list of service centres logon to
www.gemeasurement.com
Fault
Response
Power supply connected, display not lit.
Check rear panel switch set to on.
Check fuse and, if necessary, replace
Check electrical power supply fuse or circuit breaker.
24 V DC output intermittent.
Over-load internal self-resetting fuse operating.
Reduce load current to specified value.
Instruments functions, but does not reach all
set-points.
Check pneumatic supplies for correct pressures.
Check system for leaks.
In measure mode with output port sealed, the
pressure continues to increase or decrease.
Increasing pressure, leaking Apply control valve.
Decreasing pressure, leaking Release control valve.
Confirm by isolating pressure supplies.
Contact GE approved service agent.
Display pressure reading in red
Over-range, use vent de-pressurize or vent manually.
Instrument enters measure mode without user
request or command.
Idle time-out enabled but timeout period setting too
short.
Instrument will not zero.
Blocked vent port.
Check for blockage.
Contact approved service agent for repair.
Instrument controlling to set-point, no
pneumatic output
Blocked isolation valve.
Contact approved service agent for repair.
Erratic or inaccurate zero
Leaking isolation valve.
Reference port restrictor not fitted.
Contact approved service agent for repair.
Increased gas consumption.
Unstable control at set-point or does not
achieve set-point.
System leak.
Carry out leak test.
Contact approved service agent for repair.
Reference port restrictor not fitted.
If the controlled pressure stays within the
tolerance band and the pressure at the output
is within limits. If the controller status indicator
is outside the tolerance band,
Leak in the system or the supply pressure differs from the
pressure for which the control valves have been
characterised.
Содержание PACE5000
Страница 9: ...K0443 Revision A viii EN English 6 14 Vacuum System Parts 6 43 ...
Страница 49: ...4 Maintenance K0443 Revision A 4 6 EN English Intentionally Blank ...
Страница 53: ...5 Testing and Fault Finding K0443 Revision A 5 4 EN English Intentionally Blank ...
Страница 97: ...6 Reference and Specification K0443 Revision A 6 44 EN English Intentionally Blank ...