4 Command and Query Summary
K0472 Revision A
4-56
:SENS:PRES:BAR
Applicability:
PACE1000, PACE5000, PACE6000
Command Syntax
n/a
Parameter:
Short form:
Function:
Query Syntax
:SENSe[x][:PRESsure]:BARometer?
where x = 1 or 2 is the module number (default - 1)
Short form:
:SENS:BAR?
Function:
Queries the barometric pressure value.
Response:
<number> in the selected units of pressure measurement.
Description
Returns the barometric pressure measured by the optional barometric transducer. If the
optional barometric transducer is not fitted the response is zero pressure.
Example
:SENS:PRES:BAR?
:SENS:PRES:BAR 982.8430904 (this value in mbar).