PACE Series SCPI Manual
K0472 Revision A
4-17
:INST:LIM
Applicability:
PACE1000, PACE5000, PACE6000
Command Syntax
n/a
Parameter:
Short form:
Function:
Query Syntax
:INSTrument:[LIMits][x]?
where: x (sensor number) = 1, 2. . . 8 (default - 1).
Short form:
:INST?
Function:
Queries the upper and lower full-scale limits of the fitted sensor ranges.
The index number x is used to index into the list of available ranges.
Response:
A string representing the range, a number representing the upper
full-scale and a number representing the lower full-scale.
x
Sensor PACE5000, PACE6000
Sensor PACE1000
1
Module 1: control sensor
Internal sensor1
2
Module 1: source prve
Internal sensor 2
3
Module 1: source pressure -ve
Internal sensor 3
4
Module 1: barometric range (optional) External IDOS-1
5
Module 2: control sensor
External IDOS-2
6
Module 2: source prve
External IDOS-3
7
Module 2: source pressure -ve
-
8
Module 2: barometric range (optional) -