F-801 Internal RO System by GE Osmonics
12552e – 27 Feb 04
4- 29
Routine Maintenance
4.5
Product Water Culture
FREQUENCY
: Once a month or more often, depending upon culture results, and unit
protocol. We recommend before and after disinfection.
Materials Required:
•
Bucket or similar container
•
Sterile specimen container, such as Millipore (red) or 3M Commercial Standard
Plate count sampler, or a container your lab may require
RATIONALE
: This procedure will help ensure that the microbial count of the product
(RO) water falls within the AAMI Hemodialysis Water Standard. The AAMI standard
maximum allowable concentration for bacteria in product water is 200 colony-forming
units. A bacterial count above the maximum level may be harmful to the hemodialysis
patient.
OTHER
: Consult your lab to find out what type of sterile container to use and how
much product (RO) water is required to perform the test. Make certain that your
laboratory is informed of the correct procedure for performing a product water culture.
A "spread plate method" NOT a "calibrated loop" technique should be used, with
tryptic soy agar or equivalent agar, NOT blood or chocolate agar. If tested incorrectly,
the results may be inaccurate. The sample should be assayed within 30 minutes of
obtaining it, or refrigerated (5
°
C) immediately and assayed within a 24 hour period.
Obtain the samples wearing long sleeves and a mask to prevent contamination of
samples, use a “mid-stream” “clean catch” type procedure.
Содержание Osmonics F-801
Страница 2: ......
Страница 6: ...12552e D D 27 February 2004 This page intentionally left blank ...
Страница 24: ...F 801 Internal RO System by GE Osmonics 12552e 27 Feb 04 1 18 General Information NOTES ...
Страница 31: ...F 801 Internal RO System by GE Osmonics 12552e 27 Feb 04 2 7 Installation Guidelines and Instructions NOTES ...
Страница 39: ...F 801 Internal RO System by GE Osmonics 12552e 27 Feb 04 3 7 7 System Operation NOTES ...
Страница 73: ...F 801 Internal RO System by GE Osmonics 12552e 27 Feb 04 4 33 Routine Maintenance NOTES ...
Страница 87: ...F 801 Internal RO System by GE Osmonics 12552e 27 Feb 04 5 13 Troubleshooting NOTES ...
Страница 90: ...F 801 Internal RO System by GE Osmonics 12552e 27 Feb 04 6 2 Drawings This page intentionally left blank ...
Страница 100: ...F 801 Internal RO System by GE Osmonics 12552e 27 Feb 04 6 12 Drawings F 801 RO Plumbing Module Tubing ...
Страница 112: ...F 801 Internal RO System by GE Osmonics 12552e 27 Feb 04 6 24 Drawings Push In Fitting Instructions ...
Страница 113: ...F 801 Internal RO System by GE Osmonics 12552e 27 Feb 04 6 25 Drawings NOTES ...
Страница 114: ...F 801 Internal RO System by GE Osmonics 12552e 27 Feb 04 6 26 Drawings This page intentionally left blank ...
Страница 116: ...F 801 Internal RO System by GE Osmonics 12552e 27 Feb 04 7 2 Spare Parts List This page intentionally left blank ...
Страница 121: ...F 801 Internal RO System by GE Osmonics 12552e 27 Feb 04 7 7 Spare Parts List NOTES ...
Страница 122: ...F 801 Internal RO System by GE Osmonics 12552e 27 Feb 04 7 8 Spare Parts List This page intentionally left blank ...
Страница 123: ......