F-801 Internal RO System by GE Osmonics
12552e – 27 Feb 04
4- 32
Routine Maintenance
4.6.2 BIOSAN Storage Procedure
WARNING:
It is recommended that gloves (latex/vinyl) and eye protection be worn when
handling BIOSAN.
1.
Follow the BIOSAN procedure in this chapter.
2.
Do not rinse the RO out after BIOSAN has been circulated in the RO.
3.
Disconnect the disinfect line, waste line, feed water line, power and product
line.
4.
Coil the hoses and power cord, and cover the RO with a large plastic bag to
protect the RO from dust.
WARNING:
After storage, the RO unit must be completely rinsed of BIOSAN, and a bacterial
culture performed and interpreted before use. Disinfect RO unit if necessary.
Perform AAMI analysis if it has been one year or more since last done.
CAUTION:
The F-801 RO may be stored for up to two weeks in BIOSAN. After two weeks,
if longer storage is needed, the RO may be rinsed and re-packed with BIOSAN.
Содержание Osmonics F-801
Страница 2: ......
Страница 6: ...12552e D D 27 February 2004 This page intentionally left blank ...
Страница 24: ...F 801 Internal RO System by GE Osmonics 12552e 27 Feb 04 1 18 General Information NOTES ...
Страница 31: ...F 801 Internal RO System by GE Osmonics 12552e 27 Feb 04 2 7 Installation Guidelines and Instructions NOTES ...
Страница 39: ...F 801 Internal RO System by GE Osmonics 12552e 27 Feb 04 3 7 7 System Operation NOTES ...
Страница 73: ...F 801 Internal RO System by GE Osmonics 12552e 27 Feb 04 4 33 Routine Maintenance NOTES ...
Страница 87: ...F 801 Internal RO System by GE Osmonics 12552e 27 Feb 04 5 13 Troubleshooting NOTES ...
Страница 90: ...F 801 Internal RO System by GE Osmonics 12552e 27 Feb 04 6 2 Drawings This page intentionally left blank ...
Страница 100: ...F 801 Internal RO System by GE Osmonics 12552e 27 Feb 04 6 12 Drawings F 801 RO Plumbing Module Tubing ...
Страница 112: ...F 801 Internal RO System by GE Osmonics 12552e 27 Feb 04 6 24 Drawings Push In Fitting Instructions ...
Страница 113: ...F 801 Internal RO System by GE Osmonics 12552e 27 Feb 04 6 25 Drawings NOTES ...
Страница 114: ...F 801 Internal RO System by GE Osmonics 12552e 27 Feb 04 6 26 Drawings This page intentionally left blank ...
Страница 116: ...F 801 Internal RO System by GE Osmonics 12552e 27 Feb 04 7 2 Spare Parts List This page intentionally left blank ...
Страница 121: ...F 801 Internal RO System by GE Osmonics 12552e 27 Feb 04 7 7 Spare Parts List NOTES ...
Страница 122: ...F 801 Internal RO System by GE Osmonics 12552e 27 Feb 04 7 8 Spare Parts List This page intentionally left blank ...
Страница 123: ......