F-801 Internal RO System by GE Osmonics
12552e – 27 Feb 04
1 - 12
General Information
1.10 Specifications:
1.10.1 Input Water Requirements
Minimum
Maximum
Input
flow
Rate 2.0
GPM
N/A
Temperature
(water)
10°C
(50°F)
32°C
(90°F)
NOTE
:
Operating at lower temperatures may make it impossible for the F-801 RO to
produce adequate permeate for the dialysis machine.
pH (operating)
5.5
10.5
TF
Membranes
pH w/chloramines present (operating) 5.5
8.5
TF
Membranes
pH
(cleaning)
2.0
11.0
TF
Membranes
Pressure,
dynamic
30
PSI
100
PSI
1.10.2 Flow
Rates
Product Flow based on the following conditions:
New
Membranes
Operating
Pressure:
150
PSI
Product
Pressure:
1
PSI
Feed Water Temperature:
77
°
F (25
°
C)
GPM LPM GPD LPD
.41 1.55 588 2225.8
Membrane flux may vary ±15%
Waste Flow: 2000ml/min (
±
10%)
1.10.3 Electrical
Requirements
Motor
1/3 HP
Voltage 115
V~
Hertz
60
Amps
10
Phase
single
Power Cord (length)
8 ft (2.4 m)
Содержание Osmonics F-801
Страница 2: ......
Страница 6: ...12552e D D 27 February 2004 This page intentionally left blank ...
Страница 24: ...F 801 Internal RO System by GE Osmonics 12552e 27 Feb 04 1 18 General Information NOTES ...
Страница 31: ...F 801 Internal RO System by GE Osmonics 12552e 27 Feb 04 2 7 Installation Guidelines and Instructions NOTES ...
Страница 39: ...F 801 Internal RO System by GE Osmonics 12552e 27 Feb 04 3 7 7 System Operation NOTES ...
Страница 73: ...F 801 Internal RO System by GE Osmonics 12552e 27 Feb 04 4 33 Routine Maintenance NOTES ...
Страница 87: ...F 801 Internal RO System by GE Osmonics 12552e 27 Feb 04 5 13 Troubleshooting NOTES ...
Страница 90: ...F 801 Internal RO System by GE Osmonics 12552e 27 Feb 04 6 2 Drawings This page intentionally left blank ...
Страница 100: ...F 801 Internal RO System by GE Osmonics 12552e 27 Feb 04 6 12 Drawings F 801 RO Plumbing Module Tubing ...
Страница 112: ...F 801 Internal RO System by GE Osmonics 12552e 27 Feb 04 6 24 Drawings Push In Fitting Instructions ...
Страница 113: ...F 801 Internal RO System by GE Osmonics 12552e 27 Feb 04 6 25 Drawings NOTES ...
Страница 114: ...F 801 Internal RO System by GE Osmonics 12552e 27 Feb 04 6 26 Drawings This page intentionally left blank ...
Страница 116: ...F 801 Internal RO System by GE Osmonics 12552e 27 Feb 04 7 2 Spare Parts List This page intentionally left blank ...
Страница 121: ...F 801 Internal RO System by GE Osmonics 12552e 27 Feb 04 7 7 Spare Parts List NOTES ...
Страница 122: ...F 801 Internal RO System by GE Osmonics 12552e 27 Feb 04 7 8 Spare Parts List This page intentionally left blank ...
Страница 123: ......