139
Function:
1. Cabinet monitor will trip the E-Off contactor in the PDU when
a) Leak Sensor 2 detects water leakage or are open
b) T
emperature of system cabinet is over 45 degree centigrade
c) 4KW LCS or 8KW LCS/MCS stop working
2. Cabinet monitor only sends the error message to host through CAN link when
a) Leak Sensor 3 detects water leakage or are open
b) T
emperature of system cabinet is over 40 and less 45 degree centigrade
c) Water level is low in any one of LCS or MCS
3. Cabinet monitor will trip the Night Mode contactor in the PDU when
The system haven
’
t scanned for one hour.
The Nigh Mode control signal is from IRF I/O board J23 to SCIF J101, then to SCIF J105, then to J10
in cabinet monitor.
4. Cabinet monitor will
energize
the Night Mode contactor in the PDU when
The system is running TPS reset or begins to scan.
5. Provide Is24V DC power to whole CAN link.
Pump flow sensor and Tank level sensor are all Normally Closed switch, which means:
1. When LCS or MCS stop working, the pump flow sensor are close. And when they works normally,
the sensors are open.
2. When Tank level is low, the tank level sensor are close. And when the level is normal, the sensors
are open.
You can use DVM to check the resistance between the two sensor cables.
Содержание OPTIMA MR360
Страница 3: ...1...
Страница 4: ...2...
Страница 5: ...3...
Страница 7: ...5...
Страница 8: ...6...
Страница 11: ...9 The portable GasWatch2 Oxygen sensor as pictured above part number is 5111049...
Страница 14: ...12...
Страница 15: ...13...
Страница 16: ...14...
Страница 17: ...15...
Страница 18: ...16...
Страница 20: ...18...
Страница 21: ...19...
Страница 29: ...27 The trough on the newly designed Front Bridge is used to holding the cable track on the bottom of the cradle...
Страница 33: ...31...
Страница 38: ...36...
Страница 39: ...37...
Страница 41: ...39...
Страница 43: ...41...
Страница 44: ...42...
Страница 45: ...43...
Страница 47: ...45...
Страница 48: ...46...
Страница 49: ...47...
Страница 50: ...48...
Страница 51: ...49...
Страница 52: ...50...
Страница 53: ...51...
Страница 54: ...52...
Страница 55: ...53...
Страница 56: ...54...
Страница 62: ...60...
Страница 65: ...63...
Страница 66: ...64...
Страница 67: ...65...
Страница 68: ...66...
Страница 69: ...67...
Страница 71: ...69...
Страница 78: ...Image can be previewed in this new version of Image Management 76...
Страница 79: ...77...
Страница 80: ...78...
Страница 81: ...79...
Страница 82: ...80...
Страница 83: ...You can access the same diagnostic items by System Function or Hardware Location Their functions are same 81...
Страница 85: ...83...
Страница 88: ...86...
Страница 89: ...87...
Страница 90: ...88...
Страница 91: ...89...
Страница 96: ...94...
Страница 97: ...95...
Страница 102: ...100 This slide shows the detailed configuration of THTF host...
Страница 104: ...102...
Страница 105: ...103...
Страница 106: ...104...
Страница 110: ...108 MPS Magnet Power Supply is for ramping up power supply It will have no output after E stop contactor is tripped...
Страница 112: ...110...
Страница 116: ...114...
Страница 117: ...115...
Страница 124: ...122...
Страница 129: ...127...
Страница 131: ...129...
Страница 134: ...132 The part is the 200V DC power supply for three XFAs and 200V to 700V DC DC power supply...
Страница 135: ...133...
Страница 136: ...134 Three sets of DC700V and 700V are isolated...
Страница 142: ...140...
Страница 143: ...141...
Страница 147: ...145...
Страница 150: ...148...
Страница 154: ...152...
Страница 166: ...164 This table shows all the scan mode which the HNA AA and PA have and which elements will be used in different mode...
Страница 170: ...168...
Страница 171: ...169...
Страница 172: ...170...
Страница 174: ...172...
Страница 176: ...174...
Страница 180: ...178...
Страница 181: ...179 Before using the software to do the LVShim LVshim localizer scan must be done to assure the phantom is in the center...
Страница 185: ...183...
Страница 186: ...184...
Страница 187: ...185...
Страница 188: ...186 The report file is stored in export home signa bin...
Страница 189: ...187...
Страница 190: ...188...
Страница 192: ...190 The RF and DC cables for Express PA Coil Cradle mode are shipped with the system...
Страница 193: ...191...
Страница 220: ...Page 2 System Cabinet 1 2 3 4 5 6 2 3 4 5 6 1...
Страница 221: ...Page 3 System Cabinet 1 2 3 4 5 6 1 3 4 2 5 6...
Страница 222: ...Page 4 System Cabinet 1 2 3 4 5 6 4 5 6 2 3 1...
Страница 223: ...Page 5 System Cabinet 1 2 3 4 5 6 7 8 9 10 11 2 3 4 5 6 7 8 9 10 11 1...
Страница 224: ...Page 6 Magnet Side Kit 1 2 3 1 2 3...
Страница 225: ...Page 7 Rear Pedestal 1 2 3 4 5 6 7 4 3 1 2 5 6 7...
Страница 226: ...Page 8 LPCA 1 2 1 2...
Страница 227: ...Page 9 OC Computer 1 2 3 4 5 6 2 3 4 5 6 1...