112
1) The Control Board Assembly gets command signal serially from IRF I/O J23 in CAM-Lite2 cabinet. The
Control Board generates signals needed for control of the following entities:
•
The Patient Alignment Laser
•
The Bore Light
•
The Patient comfort fan relay
•
The enable signal needed by the Servo amplifier
•
The output relays in the longitudinal motor drive circuit
•
Current feedback signal for monitoring the dock motor current
•
Communicates status of power supplies and servo amplifier to the CCC board
2) The Servo Amplifier is a Pulse Width Modulated (PWM) device used for powering the longitudinal
motor. DC 42V generated by the control board is the base input voltage. The SRI send command signal
of In and Out or Fast In and Fast Out directly to Servo Amplifier to drive the longitudinal motor.
3) The CAN Core Communication (CCC) board monitors the health of the power supplies and the status
of the servo amplifier, and communicates this information to the host. The CCC board receives power
supply voltages, digital I/O signals, and analog inputs corresponding to the power supply voltages
from the Control Board. The digital I/O of the CCC operates at 3.3V so all logic levels must be converted
to this level on the Control Board. Additionally, the CCC gets an isolated 24V from an external source
for powering the serial interface communications.
3. Block diagram
Signal Table_DC_OK comes from Power Supply for PHPS Lite2. PS for PHPS Lite2 provides DC +42V and
+8V to PHPS Lite2 by J6. The signal is activated when DC +42V is OK.
Signal Mag_DC_OK comes from Power Supply for LED Power Box. PS for LED Power Box pr15V,
-15V and +24V DC power to LED Power Box and controlled DC +5V to Patient Alignment Laser and Bore
Light. The signal is activated when Power Supply works normally.
Содержание OPTIMA MR360
Страница 3: ...1...
Страница 4: ...2...
Страница 5: ...3...
Страница 7: ...5...
Страница 8: ...6...
Страница 11: ...9 The portable GasWatch2 Oxygen sensor as pictured above part number is 5111049...
Страница 14: ...12...
Страница 15: ...13...
Страница 16: ...14...
Страница 17: ...15...
Страница 18: ...16...
Страница 20: ...18...
Страница 21: ...19...
Страница 29: ...27 The trough on the newly designed Front Bridge is used to holding the cable track on the bottom of the cradle...
Страница 33: ...31...
Страница 38: ...36...
Страница 39: ...37...
Страница 41: ...39...
Страница 43: ...41...
Страница 44: ...42...
Страница 45: ...43...
Страница 47: ...45...
Страница 48: ...46...
Страница 49: ...47...
Страница 50: ...48...
Страница 51: ...49...
Страница 52: ...50...
Страница 53: ...51...
Страница 54: ...52...
Страница 55: ...53...
Страница 56: ...54...
Страница 62: ...60...
Страница 65: ...63...
Страница 66: ...64...
Страница 67: ...65...
Страница 68: ...66...
Страница 69: ...67...
Страница 71: ...69...
Страница 78: ...Image can be previewed in this new version of Image Management 76...
Страница 79: ...77...
Страница 80: ...78...
Страница 81: ...79...
Страница 82: ...80...
Страница 83: ...You can access the same diagnostic items by System Function or Hardware Location Their functions are same 81...
Страница 85: ...83...
Страница 88: ...86...
Страница 89: ...87...
Страница 90: ...88...
Страница 91: ...89...
Страница 96: ...94...
Страница 97: ...95...
Страница 102: ...100 This slide shows the detailed configuration of THTF host...
Страница 104: ...102...
Страница 105: ...103...
Страница 106: ...104...
Страница 110: ...108 MPS Magnet Power Supply is for ramping up power supply It will have no output after E stop contactor is tripped...
Страница 112: ...110...
Страница 116: ...114...
Страница 117: ...115...
Страница 124: ...122...
Страница 129: ...127...
Страница 131: ...129...
Страница 134: ...132 The part is the 200V DC power supply for three XFAs and 200V to 700V DC DC power supply...
Страница 135: ...133...
Страница 136: ...134 Three sets of DC700V and 700V are isolated...
Страница 142: ...140...
Страница 143: ...141...
Страница 147: ...145...
Страница 150: ...148...
Страница 154: ...152...
Страница 166: ...164 This table shows all the scan mode which the HNA AA and PA have and which elements will be used in different mode...
Страница 170: ...168...
Страница 171: ...169...
Страница 172: ...170...
Страница 174: ...172...
Страница 176: ...174...
Страница 180: ...178...
Страница 181: ...179 Before using the software to do the LVShim LVshim localizer scan must be done to assure the phantom is in the center...
Страница 185: ...183...
Страница 186: ...184...
Страница 187: ...185...
Страница 188: ...186 The report file is stored in export home signa bin...
Страница 189: ...187...
Страница 190: ...188...
Страница 192: ...190 The RF and DC cables for Express PA Coil Cradle mode are shipped with the system...
Страница 193: ...191...
Страница 220: ...Page 2 System Cabinet 1 2 3 4 5 6 2 3 4 5 6 1...
Страница 221: ...Page 3 System Cabinet 1 2 3 4 5 6 1 3 4 2 5 6...
Страница 222: ...Page 4 System Cabinet 1 2 3 4 5 6 4 5 6 2 3 1...
Страница 223: ...Page 5 System Cabinet 1 2 3 4 5 6 7 8 9 10 11 2 3 4 5 6 7 8 9 10 11 1...
Страница 224: ...Page 6 Magnet Side Kit 1 2 3 1 2 3...
Страница 225: ...Page 7 Rear Pedestal 1 2 3 4 5 6 7 4 3 1 2 5 6 7...
Страница 226: ...Page 8 LPCA 1 2 1 2...
Страница 227: ...Page 9 OC Computer 1 2 3 4 5 6 2 3 4 5 6 1...