Programming the 4x40 Optica
5-7
January 2006
Data Fields
Pressure Input
Table 5-7: Data Field Options
Setting
Description
Available Options
Line 1 Parameter
Choose which parameter will be
output on Line 1.
Humidity, Temperature, Pressure or
User
Line 1 Units
Set the units for the selected
parameter.
See Table 4-1 on page 4-4 for the
parameter chosen.
Line 1 Decimals
Enter the number of decimal places
for displayed data.
Enter a number (0-6).
Line 2 Parameter
Choose which parameter will be
output on Line 2.
Humidity, Temperature, Pressure or
User
Line 2 Units
Set the units for the selected
parameter.
See Table 4-1 on page 4-4 for the
parameter chosen.
Line 2 Decimals
Enter the number of decimal places
for displayed data.
Enter a number (0-6).
Line 3 Parameter
Choose which parameter will be
output on Line 3.
Humidity, Temperature, Pressure or
User
Line 3 Units
Set the units for the selected
parameter.
See Table 4-1 on page 4-4 for the
parameter chosen.
Line 3 Decimals
Enter the number of decimal places
for displayed data.
Enter a number (0-6).
Table 5-8: Pressure Input Options
Setting
Description
Available Options
Analog Input #
Choose which pressure input channel
is in use, or disable this input.
4-20 mA, 0-5 volt, User Default
Input Units
Choose the units for pressure units.
See Table 4-1 on page 4-4 for units.
Input Upper
Set the pressure that corresponds to
full-scale input.
Enter a number.
Input Lower
Set the pressure that corresponds to
zero volts or 4 mA input.
Enter a number.
Input Default
Pressure value to be used if a
measured input is disabled.
Enter a number.
Process Pressure
Default
Set whether the process pressure is
active (enabled) (see page 3-4).
Enabled/Disabled
Process Pressure
Enter the process pressure (if this
feature is enabled).
Enabled/Disabled
Содержание Optica
Страница 1: ...GE Sensing Optica General Eastern Dew Point Analyzer Operator s Manual ...
Страница 11: ...Chapter 1 ...
Страница 20: ...Chapter 2 ...
Страница 43: ...Chapter 3 ...
Страница 58: ...Chapter 4 ...
Страница 80: ...Chapter 5 ...
Страница 81: ...Programming the 4x40 Optica Introduction 5 1 Programming Technique 5 2 Programmable Functions 5 4 ...
Страница 94: ...Chapter 6 ...
Страница 95: ...Network Based Programming Introduction 6 1 Programming Screens 6 1 ...
Страница 99: ...Chapter 7 ...
Страница 109: ...Appendix A ...
Страница 114: ...Appendix B ...
Страница 115: ...Humidity Equations and Conversion Chart Introduction B 1 Vapor Pressure B 1 Humidity B 2 ...
Страница 118: ...Humidity Equations and Conversion Chart B 3 January 2006 Figure B 1 Graphical Humidity Conversion Chart ...
Страница 119: ...Appendix C ...
Страница 120: ...Configuring the Serial Interface Wiring to a Personal Computer C 1 ...
Страница 122: ...Appendix D ...
Страница 123: ...Chilled Mirror Sensors Introduction D 1 Depression D 1 Measurement Range D 2 Comparing Optica Models D 3 ...
Страница 127: ...Appendix E ...
Страница 128: ...Glossary ...
Страница 130: ...Appendix F ...
Страница 133: ...Appendix G ...
Страница 134: ...Communicating with the OPTICA Using Ethernet Direct Communication G 1 Computer Communication G 1 ...