Operation
3-5
January 2006
Scenario 1: Measurement
Without Enabling the
Process Pressure Feature
Process pressure is 100psi, within the range of a typical GE
Infrastructure Sensing chilled mirror sensor. Since vapor pressure and
dew point are pressure-dependent, a flow meter is installed
downstream of the sensor to assure that the sensor cavity is at the
process pressure (see Figure 3-5 below).
Figure 3-5: Measurement Scenario 1
Measuring Vapor Pressure
Without a Pressure Sensor
To measure vapor pressure without a pressure sensor, make the
following entries in the Pressure Input menu section for this example:
•
Input: Use Default
•
Units: psia
•
Default: 100
•
Process Status: Disabled
Measuring Vapor Pressure
With a Pressure Sensor
To measure vapor pressure with a 4-20 mA, 0-30 psia pressure sensor,
connect the sensor to the Optica’s terminal block and make the
following entries in the Pressure Input menu:
•
Input: Iin 4-20
•
Units: psia
•
Upper: 30.00
•
Lower: 0.00
•
Process Status: Disabled
Measuring Dew Point
In this scenario, dew point can be measured without knowledge of gas
pressure. Connect the equipment as shown above, with or without a
pressure sensor, and follow the normal operating procedures.
Содержание Optica
Страница 1: ...GE Sensing Optica General Eastern Dew Point Analyzer Operator s Manual ...
Страница 11: ...Chapter 1 ...
Страница 20: ...Chapter 2 ...
Страница 43: ...Chapter 3 ...
Страница 58: ...Chapter 4 ...
Страница 80: ...Chapter 5 ...
Страница 81: ...Programming the 4x40 Optica Introduction 5 1 Programming Technique 5 2 Programmable Functions 5 4 ...
Страница 94: ...Chapter 6 ...
Страница 95: ...Network Based Programming Introduction 6 1 Programming Screens 6 1 ...
Страница 99: ...Chapter 7 ...
Страница 109: ...Appendix A ...
Страница 114: ...Appendix B ...
Страница 115: ...Humidity Equations and Conversion Chart Introduction B 1 Vapor Pressure B 1 Humidity B 2 ...
Страница 118: ...Humidity Equations and Conversion Chart B 3 January 2006 Figure B 1 Graphical Humidity Conversion Chart ...
Страница 119: ...Appendix C ...
Страница 120: ...Configuring the Serial Interface Wiring to a Personal Computer C 1 ...
Страница 122: ...Appendix D ...
Страница 123: ...Chilled Mirror Sensors Introduction D 1 Depression D 1 Measurement Range D 2 Comparing Optica Models D 3 ...
Страница 127: ...Appendix E ...
Страница 128: ...Glossary ...
Страница 130: ...Appendix F ...
Страница 133: ...Appendix G ...
Страница 134: ...Communicating with the OPTICA Using Ethernet Direct Communication G 1 Computer Communication G 1 ...