Chapter 4: Installation and Setup
Step
Instructions
4. Plug the headset into the transmitter’s headset
connector
5. Rub the face of the ultrasound transducer. Verify
that you hear ultrasound audio tones from both
sides of the headset or in the internal speakers (if
head set is not connected)
Testing the ECG Functions
Perform the following steps to test the ECG functions.
Step
Instructions
Instructions to test the ECG functions:
1. Plug an ECG transducer into the ECG connector on
the transmitter as shown in the figure.
2.
Verify the following on the monitor:
•
ECG transducer connector status indication is
reflected on the transmitter display.
•
Telemetry connected symbol will be printed on
the strip chart paper. Additionally, Telemetry
symbol will be displayed on Corometrics 120,
250 and 250cx series monitor.