About this Manual
Symbol
Definition
Battery Runtime Indicator
IEC TYPE B EQUIPMENT. Type B equipment is suitable for intentional, external
and internal application to the patient, excluding direct cardiac application.
IEC TYPE BF EQUIPMENT. Type BF equipment is suitable for intentional external
and internal application to the patient, excluding direct cardiac application.
Type BF equipment has an F-type applied part.
IEC TYPE CF EQUIPMENT: Type CF equipments provide the highest degree of
patient protection. This degree of protection is achieved by increased isolation
from earthed parts and other accessible parts of the equipment, further limiting
the magnitude of possible current flow through the patient. Type CF applied
parts are suitable for direct cardiac application.
DEFIBRILLATOR-PROOF TYPE BF EQUIPMENT: Type BF equipment is suitable for
intentional external and internal application to the patient, excluding direct
cardiac appliction. Type BF equipment is type B equipment with an F-type
isolated (floating) part. The paddles indicate the equipment is defibrillator proof.
0
POWER OFF: disconnection from the mains.
I
POWER ON: connection to the mains.
1. Use Transducer in a watery environment only when connected to a telemetry
system. Do not allow the telemetry system to get wet.
2. Do not use transducer in a watery environment when directly connected to
a fetal/maternal monitor that is directly connected to AC line power.
Date of Manufacture (in “YYYY-MM” format)
Elevated, potentially hazardous levels of non-ionizing radiation.
Ultrasound
ECG