M
Millenium™ RO System by GE Osmonics
14693e - 31 Mar 03
5- 10
Troubleshooting
5.11 Low Percent Rejection
The overall membrane performance can be affected by a number of factors. Poor
rejection performance, as well as high TDS, can result from fouled membranes, low
pump pressure, inadequate waste flow, or feed water problems. An apparent
performance problem could be caused by a defective sensor or circuit board, and not be a
membrane issue at all. The recommended first step for any ‘membrane performance’
question is to confirm the rejection percentage and product TDS with a hand-held
RO/TDS meter. If the product water quality is poor, investigate the membrane.
However, if the water quality tests ‘normal’, investigate the monitoring circuits.
1.
Waste flow too low.
a. Disinfectant or cleaner not rinsed from RO.
b. Check waste flow control.
c. Check for obstruction or kink in waste line.
2.
Pump pressure too low.
a. Adjust pressure regulator.
b. Adjust pump settings.
3.
Product flow too high.
a. See
5.9
4.
Membrane scaled or fouled.
a. Verify condition of pretreatment equipment.
b. See membrane cleaning procedure in this manual.
5.
Erroneous meter readout.
a. Verify water quality with independent meter and check conductivity probe
connection.
b. Check all quick-disconnect connections.
c. Replace inlet water sensor probe.
d. Replace product water sensor probe (percent rejection).
e. Replace monitor circuit board.
6.
Membrane failure.
a. Replace
membrane.
7.
Change in feed water TDS.
a. Verify feed and product water quality with independent meter.
b. Softener problem, rinsing out high levels of salt.
Содержание MILLENIUM OSMONICS MILLENIUMTM REVERSE OSMOSIS...
Страница 1: ...GE OSMONICS MILLENIUM REVERSE OSMOSIS UNIT Operation and Maintenance Manual SUPERCEDES P N 14693 REV D...
Страница 2: ......
Страница 6: ...14693e D D 31 March 2003 This page intentionally left blank...
Страница 24: ...M Millenium RO System by GE Osmonics 14693e 31 Mar 03 1 18 General Information NOTES...
Страница 33: ...M Millenium RO System by GE Osmonics 14693e 31 Mar 03 3 2 2 System Operation This page intentionally left blank...
Страница 36: ...M Millenium RO System by GE Osmonics 14693e 31 Mar 03 3 5 5 System Operation NOTES...
Страница 37: ...M Millenium RO System by GE Osmonics 14693e 31 Mar 03 3 6 6 System Operation This page intentionally left blank...
Страница 44: ...M Millenium RO System by GE Osmonics 14693e 31 Mar 03 4 7 Routine Maintenance Tubing connections...
Страница 67: ...M Millenium RO System by GE Osmonics 14693e 31 Mar 03 4 30 Routine Maintenance NOTES...
Страница 69: ...M Millenium RO System by GE Osmonics 14693e 31 Mar 03 5 2 Troubleshooting This page intentionally left blank...
Страница 71: ...M Millenium RO System by GE Osmonics 14693e 31 Mar 03 5 4 Troubleshooting This page intentionally left blank...
Страница 80: ...M Millenium RO System by GE Osmonics 14693e 31 Mar 03 5 13 Troubleshooting NOTES...
Страница 81: ...M Millenium RO System by GE Osmonics 14693e 31 Mar 03 5 14 Troubleshooting This page intentionally left blank...
Страница 82: ...M Millenium RO System by GE Osmonics 14693e 31 Mar 03 6 1 Drawings Millenium RO System CHAPTER SIX DRAWINGS...
Страница 83: ...M Millenium RO System by GE Osmonics 14693e 31 Mar 03 6 2 Drawings This page intentionally left blank...
Страница 84: ...M Millenium RO System by GE Osmonics 14693e 31 Mar 03 6 3 Drawings Millenium RO Component Layout...
Страница 85: ...M Millenium RO System by GE Osmonics 14693e 31 Mar 03 6 4 Drawings Millenium RO Disinfect View...
Страница 86: ...M Millenium RO System by GE Osmonics 14693e 31 Mar 03 6 5 Drawings Millenium RO Flow Schematic...
Страница 96: ...M Millenium RO System by GE Osmonics 14693e 31 Mar 03 6 15 Drawings Push In Fitting Instructions...
Страница 97: ...M Millenium RO System by GE Osmonics 14693e 31 Mar 03 6 16 Drawings This page intentionally left blank...
Страница 98: ...M Millenium RO System by GE Osmonics 14693e 31 Mar 03 6 17 Drawings NOTES...
Страница 99: ...M Millenium RO System by GE Osmonics 14693e 31 Mar 03 6 18 Drawings This page intentionally left blank...
Страница 101: ...M Millenium RO System by GE Osmonics 14693e 31 Mar 03 7 2 Spare Parts List This page intentionally left blank...
Страница 105: ...M Millenium RO System by GE Osmonics 14693e 31 Mar 03 7 6 Spare Parts List NOTES...
Страница 106: ......