M
Millenium™ RO System by GE Osmonics
14693e - 31 Mar 03
5- 9
Troubleshooting
5.9
Product Flow Too High
1.
Rejection Satisfactory
A. Pump pressure too high.
a. Adjust pressure regulator setting to specifications.
B. Feedwater temperature too high.
a. Adjust temperature blending valve.
b. Correct plumbing error that allows the cold water to be heated.
C. Quick disconnect tubing connections not in correct port.
a. Verify QD connections.
D. Product recirculation check valve failed.
a. Refer to Maintenance Procedures for Check Valve Test.
2.
Rejection Unsatisfactory
A. Waste flow too low.
a. Check waste flow rate.
b. Check for blockage or kink in waste line.
c. Quick disconnect not connected completely.
B. High pH and chlorines in feed water.
a. Check feedwater for higher than 8.5 and the presence of chloramines.
b. Check pH reduction equipment if applicable.
c. Contact GE Osmonics.
C. ROMA internal seals damaged, dirty, or dislodged.
a. Disassemble ROMA, clean and inspect “O” rings and seals.
D. Membrane hydrolyzed (exposed to oxidant).
a. Verify condition of pretreatment equipment.
E. Membrane damaged (or ruptured).
a. Check for high product pressure.
b. Check for product line being back pressured.
5.10 Leaks
1. General
information: Any leak from any part of the system should be corrected as
soon as possible. Left unattended, leaks can cause secondary damage
necessitating expensive repairs.
2.
Refer to the Maintenance Procedures for Leak Repair.
Содержание MILLENIUM OSMONICS MILLENIUMTM REVERSE OSMOSIS...
Страница 1: ...GE OSMONICS MILLENIUM REVERSE OSMOSIS UNIT Operation and Maintenance Manual SUPERCEDES P N 14693 REV D...
Страница 2: ......
Страница 6: ...14693e D D 31 March 2003 This page intentionally left blank...
Страница 24: ...M Millenium RO System by GE Osmonics 14693e 31 Mar 03 1 18 General Information NOTES...
Страница 33: ...M Millenium RO System by GE Osmonics 14693e 31 Mar 03 3 2 2 System Operation This page intentionally left blank...
Страница 36: ...M Millenium RO System by GE Osmonics 14693e 31 Mar 03 3 5 5 System Operation NOTES...
Страница 37: ...M Millenium RO System by GE Osmonics 14693e 31 Mar 03 3 6 6 System Operation This page intentionally left blank...
Страница 44: ...M Millenium RO System by GE Osmonics 14693e 31 Mar 03 4 7 Routine Maintenance Tubing connections...
Страница 67: ...M Millenium RO System by GE Osmonics 14693e 31 Mar 03 4 30 Routine Maintenance NOTES...
Страница 69: ...M Millenium RO System by GE Osmonics 14693e 31 Mar 03 5 2 Troubleshooting This page intentionally left blank...
Страница 71: ...M Millenium RO System by GE Osmonics 14693e 31 Mar 03 5 4 Troubleshooting This page intentionally left blank...
Страница 80: ...M Millenium RO System by GE Osmonics 14693e 31 Mar 03 5 13 Troubleshooting NOTES...
Страница 81: ...M Millenium RO System by GE Osmonics 14693e 31 Mar 03 5 14 Troubleshooting This page intentionally left blank...
Страница 82: ...M Millenium RO System by GE Osmonics 14693e 31 Mar 03 6 1 Drawings Millenium RO System CHAPTER SIX DRAWINGS...
Страница 83: ...M Millenium RO System by GE Osmonics 14693e 31 Mar 03 6 2 Drawings This page intentionally left blank...
Страница 84: ...M Millenium RO System by GE Osmonics 14693e 31 Mar 03 6 3 Drawings Millenium RO Component Layout...
Страница 85: ...M Millenium RO System by GE Osmonics 14693e 31 Mar 03 6 4 Drawings Millenium RO Disinfect View...
Страница 86: ...M Millenium RO System by GE Osmonics 14693e 31 Mar 03 6 5 Drawings Millenium RO Flow Schematic...
Страница 96: ...M Millenium RO System by GE Osmonics 14693e 31 Mar 03 6 15 Drawings Push In Fitting Instructions...
Страница 97: ...M Millenium RO System by GE Osmonics 14693e 31 Mar 03 6 16 Drawings This page intentionally left blank...
Страница 98: ...M Millenium RO System by GE Osmonics 14693e 31 Mar 03 6 17 Drawings NOTES...
Страница 99: ...M Millenium RO System by GE Osmonics 14693e 31 Mar 03 6 18 Drawings This page intentionally left blank...
Страница 101: ...M Millenium RO System by GE Osmonics 14693e 31 Mar 03 7 2 Spare Parts List This page intentionally left blank...
Страница 105: ...M Millenium RO System by GE Osmonics 14693e 31 Mar 03 7 6 Spare Parts List NOTES...
Страница 106: ......