Revision L
MAC™ 3500 Resting ECG Analysis System
5-17
2021337-036
Maintenance:
Keyboard/Top Cover Assembly Removal & Reassembly
3. Remove the screws fastening the LVDS converter board to the display
assembly.
NOTE
Dispose of the old LVDS converter board in compliance with all applicable
local and federal laws.
4. Attach the new LVDS converter board to the display assembly using the two
screws removed in step 3.
5. Reconnect all the cables that were disconnected in step 2.
6. Reattach the display assembly to the device.
See
“Keyboard/Top Cover Assembly Reassembly”
instructions.
Keyboard/Top Cover Assembly Reassembly
Reverse the removal procedures to reassemble the keyboard/top cover assembly.
NOTE
When reconnecting the keyboard ribbon cable to its connector on the main
PCB, lift both ends of the plastic locking bar as you insert the ribbon cable.
Once the ribbon cable has been inserted, push down on the both sides of the
locking bar to secure the cable.
Also, be sure to pull the two ground cables through so they can be reconnected
to the display panel. If the unit has a KISS pump, you will also need to ensure
that the KISS pump suction tube and the KISS pump power supply cable are
pulled through as well.
97A
Содержание MAC 3500
Страница 1: ...GE Healthcare MACTM 3500 Resting ECG Analysis System Version 9D Service Manual 2021337 036 Revision L ...
Страница 9: ...Revision L MAC 3500 Resting ECG Analysis System 1 1 2021337 036 1 Introduction ...
Страница 10: ...1 2 MAC 3500 Resting ECG Analysis System Revision L 2021337 036 For your notes ...
Страница 19: ...Revision L MAC 3500 Resting ECG Analysis System 2 1 2021337 036 2 Equipment Overview ...
Страница 20: ...2 2 MAC 3500 Resting ECG Analysis System Revision L 2021337 036 For your notes ...
Страница 25: ...Revision L MAC 3500 Resting ECG Analysis System 3 1 2021337 036 3 Installation ...
Страница 26: ...3 2 MAC 3500 Resting ECG Analysis System Revision L 2021337 036 For your notes ...
Страница 40: ...3 16 MAC 3500 Resting ECG Analysis System Revision L 2021337 036 Installation General Assembly For your notes ...
Страница 41: ...Revision L MAC 3500 Resting ECG Analysis System 4 1 2021337 036 4 Troubleshooting ...
Страница 42: ...4 2 MAC 3500 Resting ECG Analysis System Revision L 2021337 036 For your notes ...
Страница 44: ...4 4 MAC 3500 Resting ECG Analysis System Revision L 2021337 036 Troubleshooting Assembly Descriptions ...
Страница 83: ...Revision L MAC 3500 Resting ECG Analysis System 5 1 2021337 036 5 Maintenance ...
Страница 84: ...5 2 MAC 3500 Resting ECG Analysis System Revision L 2021337 036 For your notes ...
Страница 122: ...5 40 MAC 3500 Resting ECG Analysis System Revision L 2021337 036 Maintenance Functional Checkout Procedures ...
Страница 123: ...Revision L MAC 3500 Resting ECG Analysis System 6 1 2021337 036 6 Parts Lists ...
Страница 124: ...6 2 MAC 3500 Resting ECG Analysis System Revision L 2021337 036 For your notes ...
Страница 152: ...6 30 MAC 3500 Resting ECG Analysis System Revision L 2021337 036 Parts Lists Field Replaceable Units FRUs ...
Страница 153: ...Revision L MAC 3500 Resting ECG Analysis System A 1 2021337 036 A Technical Specifications ...
Страница 154: ...A 2 MAC 3500 Resting ECG Analysis System Revision L 2021337 036 For your notes ...
Страница 160: ...A 8 MAC 3500 Resting ECG Analysis System Revision L 2021337 036 Technical Specifications Barcode Scanner For your notes ...
Страница 161: ...Revision L MAC 3500 Resting ECG Analysis System B 1 2021337 036 B Software Hardware Compatibility ...
Страница 162: ...B 2 MAC 3500 Resting ECG Analysis System Revision L 2021337 036 For your notes ...
Страница 167: ...Revision L MAC 3500 Resting ECG Analysis System C 1 2021337 036 C Electromagnetic Compatibility ...
Страница 168: ...C 2 MAC 3500 Resting ECG Analysis System Revision L 2021337 036 For your notes ...
Страница 175: ...Revision L MAC 3500 Resting ECG Analysis System Index 1 2021337 036 Index ...
Страница 176: ...Index 2 MAC 3500 Resting ECG Analysis System Revision L 2021337 036 For your notes ...
Страница 178: ...Index Index 4 MAC 3500 Resting ECG Analysis System Revision L 2021337 036 ...
Страница 179: ......