MAC™ 3500 Resting ECG Analysis System
Revision L
2021337-036
Electromagnetic Compatibility:
Electromagnetic Compatibility (EMC)
Guidance and Manufacturer’s Declaration - Electromagnetic
Immunity
The MAC 3500 system is intended for use in the electromagnetic environment
specified below. It is the responsibility of the customer or user to ensure that the
MAC 3500 system is used in such an environment.
NOTE
Ut
is the AC mains voltage prior to application of the test level.
Immunity Test
EN 60601 Test Level
Compliance Level
Electromagnetic Environment - Guidance
Electrostatic discharge
(ESD)
EN 61000-4-2
± 6 kV contact
± 8 kV air
± 6 kV contact
± 8 kV air
Floors should be wood, concrete or ceramic
tile. If floors are covered with synthetic
material, the relative humidity should be at
least 30%.
Electrical fast
transient/burst
EN 61000-4-4
± 2 kV for power supply lines
±1 kV for input/output lines
± 2 kV for power supply lines
±1 kV for input/output lines
Mains power should be that of a typical
commercial or hospital environment.
Surge
EN 61000-4-5
± 1 kV differential mode
± 2 kV common mode
± 1 kV differential mode
± 2 kV common mode
Mains power should be that of a typical
commercial or hospital environment.
Voltage dips, short
interruptions and
voltage variations on
power supply input
lines
EN 61000-4-11
<5% Ut (>95% dip in U
t
)
for 0.5 cycles
40% Ut (60% dip in U
t
)
for 5 cycles
70% Ut (30% dip in U
t
)
for 25 cycles
<5% Ut (>95% dip in U
t
)
for 5 sec
<5% Ut (>95% dip in U
t
)
for 0.5 cycles
40% Ut (60% dip in U
t
)
for 5 cycles
70% Ut (30% dip in U
t
)
for 25 cycles
<5% Ut (>95% dip in U
t
)
for 5 sec
Mains power should be that of a typical
commercial or hospital environment. If the
user of the MAC 3500 requires continued
operation during power mains interruptions,
it is recommended that the MAC 3500 be
powered from an uninterruptible power
supply or a battery.
Power frequency
(50/60 Hz) magnetic
field
EN 61000-4-8
3 A/m
3 A/m
Power frequency magnetic fields should be
at levels characteristics of a typical location
in a typical commercial or hospital
environment.
Содержание MAC 3500
Страница 1: ...GE Healthcare MACTM 3500 Resting ECG Analysis System Version 9D Service Manual 2021337 036 Revision L ...
Страница 9: ...Revision L MAC 3500 Resting ECG Analysis System 1 1 2021337 036 1 Introduction ...
Страница 10: ...1 2 MAC 3500 Resting ECG Analysis System Revision L 2021337 036 For your notes ...
Страница 19: ...Revision L MAC 3500 Resting ECG Analysis System 2 1 2021337 036 2 Equipment Overview ...
Страница 20: ...2 2 MAC 3500 Resting ECG Analysis System Revision L 2021337 036 For your notes ...
Страница 25: ...Revision L MAC 3500 Resting ECG Analysis System 3 1 2021337 036 3 Installation ...
Страница 26: ...3 2 MAC 3500 Resting ECG Analysis System Revision L 2021337 036 For your notes ...
Страница 40: ...3 16 MAC 3500 Resting ECG Analysis System Revision L 2021337 036 Installation General Assembly For your notes ...
Страница 41: ...Revision L MAC 3500 Resting ECG Analysis System 4 1 2021337 036 4 Troubleshooting ...
Страница 42: ...4 2 MAC 3500 Resting ECG Analysis System Revision L 2021337 036 For your notes ...
Страница 44: ...4 4 MAC 3500 Resting ECG Analysis System Revision L 2021337 036 Troubleshooting Assembly Descriptions ...
Страница 83: ...Revision L MAC 3500 Resting ECG Analysis System 5 1 2021337 036 5 Maintenance ...
Страница 84: ...5 2 MAC 3500 Resting ECG Analysis System Revision L 2021337 036 For your notes ...
Страница 122: ...5 40 MAC 3500 Resting ECG Analysis System Revision L 2021337 036 Maintenance Functional Checkout Procedures ...
Страница 123: ...Revision L MAC 3500 Resting ECG Analysis System 6 1 2021337 036 6 Parts Lists ...
Страница 124: ...6 2 MAC 3500 Resting ECG Analysis System Revision L 2021337 036 For your notes ...
Страница 152: ...6 30 MAC 3500 Resting ECG Analysis System Revision L 2021337 036 Parts Lists Field Replaceable Units FRUs ...
Страница 153: ...Revision L MAC 3500 Resting ECG Analysis System A 1 2021337 036 A Technical Specifications ...
Страница 154: ...A 2 MAC 3500 Resting ECG Analysis System Revision L 2021337 036 For your notes ...
Страница 160: ...A 8 MAC 3500 Resting ECG Analysis System Revision L 2021337 036 Technical Specifications Barcode Scanner For your notes ...
Страница 161: ...Revision L MAC 3500 Resting ECG Analysis System B 1 2021337 036 B Software Hardware Compatibility ...
Страница 162: ...B 2 MAC 3500 Resting ECG Analysis System Revision L 2021337 036 For your notes ...
Страница 167: ...Revision L MAC 3500 Resting ECG Analysis System C 1 2021337 036 C Electromagnetic Compatibility ...
Страница 168: ...C 2 MAC 3500 Resting ECG Analysis System Revision L 2021337 036 For your notes ...
Страница 175: ...Revision L MAC 3500 Resting ECG Analysis System Index 1 2021337 036 Index ...
Страница 176: ...Index 2 MAC 3500 Resting ECG Analysis System Revision L 2021337 036 For your notes ...
Страница 178: ...Index Index 4 MAC 3500 Resting ECG Analysis System Revision L 2021337 036 ...
Страница 179: ......