4 - 18
Section 4-3 - General Procedure
4-3-8-2
Doppler Mode Controls
Table 4-4
Doppler Mode Controls
Control
Possible Bioeffect
Description/Benefit
Auto Spectral
Optimize [ASO]
(Auto)
Yes
Auto in Doppler Mode optimizes the spectral data. Auto
adjusts the Velocity Scale/PRF (on live images only),
baseline shift, Dynamic Range and invert (if preset).
The benefit of Auto can be found in reduced
optimization time and a more consistent and accurate
optimization process.
Update
Yes
Toggles between simultaneous and update
presentation while viewing the timeline.
Doppler sample
volume gate position
(Trackball)
Yes
Moves the sample volume gate on the B-Mode's
Doppler Mode cursor. The gate is positioned over a
specific position within the vessel.
Doppler sample
volume length
Yes
Sizes the sample volume gate.
Scale (Velocity
Scale)
Yes
Adjusts the velocity scale to accommodate faster/
slower blood flow velocities. Velocity scale determines
pulse repetition frequency.
If the sample volume gate range exceeds single gate
Scale capability, the system automatically switches to
high PRF mode. Multiple gates appear, and HPRF is
indicated on the display.
Angle Correct
No
Estimates the flow velocity in a direction at an angle to
the Doppler vector by computing the angle between the
Doppler vector and the flow to be measured.
NOTE: When the Doppler Mode Cursor and angle
correct indicator are aligned (the angle is O), you
cannot see the angle correct indicator.
Quick Angle
No
Quickly adjusts the angle by 60 degrees.
Wall Filter
No
Insulates the Doppler signal from excessive noise
caused from vessel movement.
Baseline
No
Adjusts the baseline to accommodate faster or slower
blood flows to eliminate aliasing.
Mode Cursor
No
Displays the Doppler Mode cursor on the B-Mode
image.
Steer and Fine Steer Yes
You can slant the ROI of the Color Flow linear image
left or right to get more information without moving the
probe. The angle steer function only applies to linear
probes.
Volume
No
Controls audio output.
Invert
No
Vertically inverts the spectral trace without affecting the
baseline position.
Compression
No
Dynamic range controls how echo intensities are
converted to shades of gray, thereby increasing the
range of contrast you can adjust.
Trace Method
No
Traces the average mean and peak velocities in
realtime or frozen images.
Cycles to Average
No
The average value over a number of cycles (from 1-5).
Trace Sensitivity
No
Adjust the trace to follow the waveform for signal
strength.
Trace Direction
No
Specifies trace direction.
Содержание LOGIQ F Series
Страница 1: ...GE LOGIQ F Series Basic Service Manual Part Number 5446617 100 Revision 10...
Страница 2: ......
Страница 3: ...GE LOGIQ F SERIES DIRECTION 5446617 100 REVISION 10 BASIC SERVICE MANUAL i Important Precautions...
Страница 4: ...GE LOGIQ F SERIES DIRECTION 5446617 100 REVISION 10 BASIC SERVICE MANUAL ii...
Страница 5: ...GE LOGIQ F SERIES DIRECTION 5446617 100 REVISION 10 BASIC SERVICE MANUAL iii...
Страница 6: ...GE LOGIQ F SERIES DIRECTION 5446617 100 REVISION 10 BASIC SERVICE MANUAL iv...
Страница 7: ...GE LOGIQ F SERIES DIRECTION 5446617 100 REVISION 10 BASIC SERVICE MANUAL v...
Страница 8: ...GE LOGIQ F SERIES DIRECTION 5446617 100 REVISION 10 BASIC SERVICE MANUAL vi...
Страница 9: ...GE LOGIQ F SERIES DIRECTION 5446617 100 REVISION 10 BASIC SERVICE MANUAL vii...
Страница 10: ...GE LOGIQ F SERIES DIRECTION 5446617 100 REVISION 10 BASIC SERVICE MANUAL viii...
Страница 11: ...GE LOGIQ F SERIES DIRECTION 5446617 100 REVISION 10 BASIC SERVICE MANUAL ix...
Страница 12: ...GE LOGIQ F SERIES DIRECTION 5446617 100 REVISION 10 BASIC SERVICE MANUAL x...
Страница 14: ...GE LOGIQ F SERIES DIRECTION 5446617 100 REVISION 10 BASIC SERVICE MANUAL xii...
Страница 167: ...2013 2017 General Electric Company GE GEMedical Systems Ultrasound 9900 Innovation Drive Wawautosa Wisconsin 53226 USA...