Chapter 3 System Setup

3 - 7

3-3-1

Moving into Position

In general, a single adult can move the LOGIQ F Series along an even surface with no steep grades. 
At least two people should move the machine when large humps, grooves, or grades will be 
encountered. (It is better to pull from the rear rather than push from the front of the unit). Before moving, 
store all loose parts in the unit. Wrap transducers in soft cloth or foam to prevent damage.

Although LOGIQ F Series is a mobile machine, two people should move it over rough surfaces or up 
and down grades.

3-3-2

Product Locator Installation Card

NOTE:

The Product Locator Installation Card shown may not be same as the provided Product Locator 
card.

CAUTION

Do not tilt the unit more than 5 degrees to avoid tipping it over.
To avoid injury by tipping over. Set the monitor to the lowest position before moving.

Figure 3-3   Product Locator Installation Card

Содержание LOGIQ F Series

Страница 1: ...GE LOGIQ F Series Basic Service Manual Part Number 5446617 100 Revision 10...

Страница 2: ......

Страница 3: ...GE LOGIQ F SERIES DIRECTION 5446617 100 REVISION 10 BASIC SERVICE MANUAL i Important Precautions...

Страница 4: ...GE LOGIQ F SERIES DIRECTION 5446617 100 REVISION 10 BASIC SERVICE MANUAL ii...

Страница 5: ...GE LOGIQ F SERIES DIRECTION 5446617 100 REVISION 10 BASIC SERVICE MANUAL iii...

Страница 6: ...GE LOGIQ F SERIES DIRECTION 5446617 100 REVISION 10 BASIC SERVICE MANUAL iv...

Страница 7: ...GE LOGIQ F SERIES DIRECTION 5446617 100 REVISION 10 BASIC SERVICE MANUAL v...

Страница 8: ...GE LOGIQ F SERIES DIRECTION 5446617 100 REVISION 10 BASIC SERVICE MANUAL vi...

Страница 9: ...GE LOGIQ F SERIES DIRECTION 5446617 100 REVISION 10 BASIC SERVICE MANUAL vii...

Страница 10: ...GE LOGIQ F SERIES DIRECTION 5446617 100 REVISION 10 BASIC SERVICE MANUAL viii...

Страница 11: ...GE LOGIQ F SERIES DIRECTION 5446617 100 REVISION 10 BASIC SERVICE MANUAL ix...

Страница 12: ...GE LOGIQ F SERIES DIRECTION 5446617 100 REVISION 10 BASIC SERVICE MANUAL x...

Страница 13: ...comply with the requirements of the applicable electrical codes The purchaser of GE equipment shall only utilize qualified personnel i e GE s field engineers personnel of third party service companies...

Страница 14: ...GE LOGIQ F SERIES DIRECTION 5446617 100 REVISION 10 BASIC SERVICE MANUAL xii...

Страница 15: ...basic service software does not extend to or cover this software or related documentation If you are a GE employee or a customer who has entered into such a license agreement with GE to use this propr...

Страница 16: ...s and spare part 10 2017 09 25 Update the unpacking procedure of the equipment for sea transportation Pages Revision Pages Revision Title Page N A Chapter 5 Theory pages 5 1 to 5 4 10 Important Precau...

Страница 17: ...ventions 1 6 Conventions Used in Book 1 6 Standard Hazard Icons 1 7 Product Icons 1 7 Safety Considerations 1 8 Introduction 1 8 Human Safety 1 8 Mechanical Safety 1 8 Electrical Safety 1 9 Label Loca...

Страница 18: ...chapter 2 2 1 General Console Requirements 2 2 Console Environmental Requirements 2 2 Electrical Requirements 2 2 EMI Limitations 2 4 Scan Probe Environmental Requirements 2 5 Facility Needs 2 6 Purch...

Страница 19: ...r Order 3 8 Physical Inspection 3 8 EMI Protection 3 8 Completing the Installation 3 9 Power On Boot Up 3 9 Power Off Shutdown 3 9 Transducer Connection 3 10 System Configuration 3 11 System Specifica...

Страница 20: ...On Boot Up 4 2 Power Off Shutdown 4 3 Archiving and Loading Presets 4 5 Adjusting the Display Monitor 4 6 System Features 4 7 B Mode Checks 4 11 M Mode Controls 4 15 Doppler Mode Checks 4 17 Color Flo...

Страница 21: ...0 REVISION 10 BASIC SERVICE MANUAL Table of Contents 5 CHAPTER 5 Components and Functions Theory Overview 5 1 Purpose of Chapter 5 5 1 Block Diagram 5 2 System Diagram 5 2 Software Diagram 5 3 Common...

Страница 22: ...LOGIQ F SERIES DIRECTION 5446617 100 REVISION 10 BASIC SERVICE MANUAL 6 Table of Contents CHAPTER 6 Service Adjustments Overview 6 1 Purpose of this chapter 6 6 1 Monitor Adjustments 6 2 Adjustments P...

Страница 23: ...on 7 2 Collect a Trouble Image with Logs 7 3 Screen Captures 7 4 Check and Record the Store Key Function 7 4 Setting the Print Key to Screen Capture 7 4 Capturing a Screen 7 4 Reset the Store Key to C...

Страница 24: ...8 Table of Contents CHAPTER 8 Replacement Procedures Overview 8 1 Purpose of Chapter 8 8 1 DISASSEMBLY RE ASSEMBLY 8 2 Warning and Caution 8 2 Returning Shipping for repairs 8 2 Air filter 8 3 Trackb...

Страница 25: ...10 BASIC SERVICE MANUAL Table of Contents 9 CHAPTER 9 Renewal Parts Overview 9 1 Purpose of Chapter 9 9 1 List of Abbreviations 9 1 Renewal Parts Lists 9 2 Equipment Models Covered in this Chapter 9 2...

Страница 26: ...10 2 How often should care maintenance tasks be performed 10 2 Tools Required 10 4 Standard GE Tool Kit 10 4 Special Tools Supplies and Equipment 10 6 Input Power 10 7 Cleaning 10 7 Physical Inspecti...

Страница 27: ...recommended as part of the installation or as required during servicing and periodic maintenance 5 Chapter 5 Components and Functions Theory Contains block diagrams and functional explanations of the...

Страница 28: ...l Designations Part Number Description 5478043 LOGIQ F3 Console for China 5478051 LOGIQ F3 Console except China Table 1 3 LOGIQ F5 Model Designations Part Number Description 5478049 LOGIQ F5 Console f...

Страница 29: ...P D711MD 5449734 1 Y 8003 HP Officejet 100 Chinese kit 5426594 1 Y 8003A HP Officejet 100 European kit 5426595 1 Y 8003B HP Officejet 100 Japanese kit 5426596 1 Y 8003C HP Officejet 100 USA kit 542659...

Страница 30: ...biopsy kit E8385MJ 1 Y 8017 E8C RS reusable biopsy kit 2398164 1 Y 8019 L6 12 RS reusable biopsy kit 5176499 1 Y 8020 3Sc RS reusable biopsy kit 5329137 1 Y 8021 RAB2 6 RS reusable biopsy kit KTD10623...

Страница 31: ...oftware USB 5495846 4S 5495846 5S 1 Y 8022E LOGIQ F Series R1 0 4 System and Application Software USB 5495846 5S 5495846 6S 1 Y 8022F LOGIQ F Series R1 0 5 System and Application Software USB 5495846...

Страница 32: ...r potential hazards are labeled in one of following ways NOTE Notes provide important information about an item or a procedure Information contained in a NOTE can often save you time or effort DANGER...

Страница 33: ...be used to make you aware of specific types of hazards that could cause harm Other hazard icons make you aware of specific procedures that should be followed 1 2 3 Product Icons Please refer to User M...

Страница 34: ...T MAY BECOME UNSTABLE AND TIP OVER WARNING ULTRASOUND PROBES ARE HIGHLY SENSITIVE MEDICAL INSTRUMENTS THAT CAN EASILY BE DAMAGED BY IMPROPER HANDLING USE CARE WHEN HANDLING AND PROTECT FROM DAMAGE WHE...

Страница 35: ...st be connected to an electrical ground The system is equipped with a three conductor AC power cable This must be plugged into an approved electrical outlet with safety ground The power outlet used fo...

Страница 36: ...ultrasound probe The purpose of the regulation is to protect employees in the transportation industry as well as the people who will receive or open this package NOTE The USER SERVICE staff should di...

Страница 37: ...signal cables In addition to electromagnetic energy EMC also includes possible effects from electrical fields magnetic fields electrostatic discharge and disturbances in the electrical power supply 1...

Страница 38: ...umstance should a part or equipment with visible body fluids be taken or shipped from a clinic or site for example body coils or an ultrasound probe The purpose of the regulation is to protect employe...

Страница 39: ...Customer Assistance Location Phone Number USA GE Vingmed Ultrasound Ultrasound Service Engineering 9900 Innovation Drive Wauwatosa WI 53226 Service On site Service Parts Application Support 1 800 437...

Страница 40: ...anufacturer FAX Number GE Medical Systems China Co Ltd No 19 Changjiang Road Wuxi National Hi Tech Development Zone Jiangsu P R China 214028 TEL 86 510 85225888 FAX 86 510 85226688 Table 1 12 Factory...

Страница 41: ...of chapter 2 This chapter provides the information required to plan and prepare for the installation of LOGIQ F Series Included are descriptions of the facility and electrical needs to be met by the p...

Страница 42: ...asound equipment This dedicated power shall originate at the last distribution panel before the system Sites with a mains power system with defined Neutral and Live The dedicated line shall consist of...

Страница 43: ...des 2 2 2 5 Unit Power Plug If the unit arrives without the power plug or with the wrong plug you must contact your GE dealer or the installation engineer must supply what is locally required 2 2 2 6...

Страница 44: ...nding is the most likely reason a unit will have noisy images Check grounding of the power cord and power outlet Replace all screws RF gaskets covers cores After you finish repairing or updating the s...

Страница 45: ...mental Requirements Operation 3 to 40 C Storage 5 to 50 C NOTE Temperature in degrees C Conversion to Degrees F Degrees C 9 5 32 NOTICE SYSTEMS AND ELECTRONIC PROBES ARE DESIGNED FOR STORAGE TEMPERATU...

Страница 46: ...connections between pieces of electrical equipment products involved and the accompanying electrical installations are highly sophisticated and special engineering competence is required All electrica...

Страница 47: ...out a defined Neutral The dedicated line shall consist of one phase two lines not shared with any other circuit and a full size ground wire from the distribution panel to the Ultrasound outlet Please...

Страница 48: ...ICATED POWER RECEPTACLE EXTERNAL PERIPHERALS SINK COUNTER TOP FILM VIEWER FILM SUPPLIES SINK LINEN SUPPLY PROBES SUPPLIES 24 IN 61 CM EXAMINARION TABLE 76 IN 193 CM DOOR 30 IN 76 CM An 8 by 10 foot Mi...

Страница 49: ...p the on board monitor and peripherals enabling viewing to be done while scanning continues With DICOM images can be archived stored and retrieved faster easier and at a lower cost 2 3 4 4 DICOM Optio...

Страница 50: ...d Figure 2 2 Worksheet for DICOM Network Information LOGIQ F Host Name AE Title Local Port IP Address Net Mask ROUTING INFORMATION ROUTER2 ROUTER3 ROUTER1 Destination IP Addresses GATEWAY IP Addresses...

Страница 51: ...ow to check and test the unit and external peripherals for electrical safety are included in this procedure Also included LOGIQ F Series in this section are guidelines for transporting the unit to a n...

Страница 52: ...talled and checked out by an experienced service technician in approximately 1 hour LOGIQ F Series consoles with optional equipment may take slightly longer Table 3 2 Average Installation Time Descrip...

Страница 53: ...rsonnel should carry out servicing and troubleshooting NOTE For information regarding packing labels refer to LABELS ON PACKAGE 3 After being transported the unit may be very cold or hot If this is th...

Страница 54: ...onnect the unit to a properly grounded power outlet Do not use a three to two prong adapter This defeats safety grounding CAUTION Do NOT wear the ESD wrist strap when you work on live circuits and mor...

Страница 55: ...Remove the clear plastic wrapped around the system and aluminum foil bag from the unit NOTE There is no clear plastic if the system is transported by sea Ignore this step if there is no clear plastic...

Страница 56: ...GE LOGIQ F SERIES DIRECTION 5446617 100 REVISION 10 BASIC SERVICE MANUAL 3 6 Section 3 3 Receiving and Unpacking the Equipment Figure 3 2 Open the box...

Страница 57: ...e front of the unit Before moving store all loose parts in the unit Wrap transducers in soft cloth or foam to prevent damage Although LOGIQ F Series is a mobile machine two people should move it over...

Страница 58: ...GIQ F Series is found on a label to the right of the Power switch and External I O on the rear of the system 3 4 2 2 Video Formats Check that the video format is set to the locally used video standard...

Страница 59: ...herals and the Back end Processor 3 5 1 2 Turn on the system Press the Power On Off switch at the front of the system once When the Power On Off switch on the Control Panel is pressed once the Back en...

Страница 60: ...en the power status LED is turned blue 5 Disconnect the probes Clean or disinfect all probes as necessary Store them in their shipping cases to avoid damage 3 5 2 2 Scanner Shutdown Disconnect the Mai...

Страница 61: ...6 System Configuration 3 6 1 System Specifications 3 6 1 1 Physical Dimensions The physical dimensions of the LOGIQ F Series unit are summarized in Table 3 4 on page 3 11 The Size of LOGIQ F Series Ta...

Страница 62: ...Current Frequency LOGIQ F Series 100 240 V AC 400VA 50 60Hz Table 3 7 List of Optional Peripherals B W Printer SONY UP D897MD USB B W Printer SONY UP D898MD USB B W Printer SONY UP D711MD USB HP Print...

Страница 63: ...AC printer can be connected to the isolated USB port on the rear panel only NOTE There is only one isolated USB port on the system as shown above Figure 3 6 Rear Connector Panel Table 3 8 Peripheral...

Страница 64: ...13 HSY 6 GND 14 VSY 7 GND 15 NC 8 GND Table 3 10 Pin Assignments of USB Pin No Signal Pin No Signal 1 5 VDC 5 5 VDC 2 DATA 6 DATA 3 DATA 7 DATA 4 GND 8 GND Table 3 11 Pin Assignments of Audio Pin No...

Страница 65: ...ected using USB Ports Sony UP D711MD Printer is connected to the LOGIQ F Series system via the USB port on the USB hub under the control panel NOTE The DC printer can only be connected to the system v...

Страница 66: ...be properly connected using the isolated USB Port And connect the power cable of UP D25MD to the wall outlet C Connect the B W printer Sony UP 897 UP D898MD to the system The B W printer can be prope...

Страница 67: ...ipherals cont d D Connect Foot Switch to the system Foot Switch can be properly connected using USB Ports E Connect the USB Memory to the system The USB Memory can be properly connected using USB port...

Страница 68: ...nect HP Officejet 100 printer to the system HP Officejet 100 Printer can be properly connected using the isolated USB Port And connect the power cable of HP Officejet 100 to the wall outlet H Connect...

Страница 69: ...Print in the Service list Select the printer from the Printer pull down Properties menu For the UP D897 printer select Portrait as orientation Type the printer name in the Name field This name is used...

Страница 70: ...Sony UP D897 as an example 1 Press Utility System Peripherals Select the UP D897 from the pull down menu under Standard Printer Properties Click Properties 2 Select Properties from Printer pull down...

Страница 71: ...r to the LOGIQ F Series Basic User Manual Chapter 16 Customizing Your System For information on configuring DICOM Connectivity Refer to the LOGIQ F Series Basic User Manual Chapter 16 Customizing Your...

Страница 72: ...Site System Information Services Destination Devices TCP IP Settings Device Type Manufacturer Name IP Address Port AE Title 1 2 3 4 5 6 7 8 9 10 11 12 Name AE Title Dept LOGIQ SN Type Floor Room REV C...

Страница 73: ...E LOGIQ F SERIES DIRECTION 5273008 100 REVISION 10 BASIC SERVICE MANUAL Chapter 3 System Setup 3 23 Section 3 9 Loading Base Image Software Refer to Section 8 5 Loading Base Image Software on page 8 7...

Страница 74: ...are Version check out Section 3 10Software Version check out 3 10 1 Functional Check out 1 Power on LOGIQ F Series scanner and wait until system booting to main screen 2 Select Utility on control pane...

Страница 75: ...ety and user information is available during the operation and service of the complete system 3 11 1 Product Locator Installation NOTE The Product Locator Installation Card shown may not be same as th...

Страница 76: ...GE LOGIQ F SERIES DIRECTION 5446617 100 REVISION 10 BASIC SERVICE MANUAL 3 26 Section 3 11 Paperwork This page was intentionally left blank...

Страница 77: ...F Series console diagnostics by using the built in service software and power supply Section 4 2 Required Equipment To perform these tests you ll need any of the sector linear or convex transducers n...

Страница 78: ...e Cooling Unit Control Panel LCD Peripherals and the Back end Processor 4 3 1 2 Turn on the system Press the Power On Off switch at the front of the system once When the Power On Off switch on the Con...

Страница 79: ...ystem Exit window is displayed 3 Using the Trackball or Select key select Shutdown 4 The shutdown process takes a few seconds and the power off sequence is complete when the power status LED is turned...

Страница 80: ...has been changed by 24 hours earlier than the current system time of last boot up This warning message is to remind the user to check the system date in case the system date and time is incorrect Move...

Страница 81: ...rchiving Presets to an CD R Disk or USB memory device 1 Insert an empty blank CD R disk into the DVD RW 2 Access the Utility Menu and select System The Backup sheet will be shown on the LCD display 3...

Страница 82: ...he DVD RW 2 Access to the Utility Menu and select System The Restore sheet will be shown on the LCD display 3 Select the item to restore either from Resource Files 4 Enter restore destination or brows...

Страница 83: ...itor Display Figure 4 5 Control Panel Tour 1 Power On Off 2 Page Up Down keys 3 Rotary Button 4 TGC 5 User Defined keys 6 Patient key 7 Probe Preset keys 8 Worksheet key 9 End Exam key 10 Archive key...

Страница 84: ...hioces 4 End Exam Activates Image Management and Touch panel with end of exam options 5 Utility Activates system configuration menus 6 Model Selects the application to use 7 Probe Indicator Indicates...

Страница 85: ...general the key name is indicated at the top of the key There are different types of Touch panel keys as illustrated below 1 Press to toggle control on off 2 Progress Select keys are used for control...

Страница 86: ...en 8 Delete Image 9 Save As Menu 10 Number of Images in Exam 11 Page Indication 12 Trackball Functionality Status 13 Current date and time Caps Lock lit when on network connection indicator PC connect...

Страница 87: ...1 Preparations 1 Connect one of the probes listed in 3 6 6 Available Probes on page 3 21 in Chapter 3 Installation to the System probe connector 2 Turn ON the scanner if it isn t turned on already Fi...

Страница 88: ...om and in Spectral Doppler Auto in PW Doppler Mode ASO optimizes the spectral data Auto adjusts the Velocity Scale PRF live imaging only baseline shift dynamic range and invert if preset Upon deactiva...

Страница 89: ...intensity valuations Colorize is NOT a Doppler Mode NOTE You can colorize realtime or CINE images or Timeline CINE Colorizes the gray scale image to enhance the eye s discrimination capability Spectr...

Страница 90: ...ailable in 2D imaging and may be used with any transducer or clinical application when image speckle appears to interfere with the desired image detail LOGIQView Option No LOGIQView provides the abili...

Страница 91: ...s 4 3 7 1 Preparations 1 Connect one of the probes listed in 3 6 6 Available Probes on page 3 21 in Chapter 3 Installation to the System probe connector 2 Turn ON the scanner if it isn t turned on alr...

Страница 92: ...Control Possible Bioeffect Description Benefit Sweep Speed No Changes the speed at which the timeline is swept Available in M Mode Doppler Mode and M Color Flow Mode Anatomical M Mode option Yes Anato...

Страница 93: ...rations 1 Connect one of the probes listed in 3 6 6 Available Probes on page 3 21 in Chapter 3 Installation to the System probe connector 2 Turn ON the scanner if it isn t turned on already Figure 4 1...

Страница 94: ...the display Angle Correct No Estimates the flow velocity in a direction at an angle to the Doppler vector by computing the angle between the Doppler vector and the flow to be measured NOTE When the D...

Страница 95: ...calculations are automatically calculated Auto Calcs No Activates the calculation automatically which you select in the Modify Auto Calculation when the system is in a state of freeze or live Simulta...

Страница 96: ...nly support by LOGIQ C5 Premium 4 3 9 1 Preparations 1 Connect one of the probes listed in 3 6 6 Available Probes on page 3 21 in Chapter 3 Installation to the System probe connector 2 Turn ON the sca...

Страница 97: ...ect to reverse flow or vice versa Baseline adjusts the alias point The default baseline is at the midpoint of the color display and at the midpoint of the color bar reference display Angle Steer Yes Y...

Страница 98: ...based on the number of reflectors that are moving regardless of their velocity PDI does not map velocity therefore it is not subject to aliasing TVI Option Yes Tissue Velocity Imaging TVI calculates a...

Страница 99: ...adjust Cursor Select NOTE After you complete the measurement to erase all data that has been measured to this point but not data entered onto worksheets select Clear 4 3 10 2 Circumference Area Ellip...

Страница 100: ...ly position the probe cord so it is free to move and is not resting on the floor 4 3 11 2 Activating the probe The probe activates in the currently selected operating mode The probe s default settings...

Страница 101: ...of the CINE Loop Adjust the Start Frame up down Two Button Softkey upward to move forward through the CINE Loop Adjust the Softkey downward to move backward through the CINE Loop Press the End Frame T...

Страница 102: ...type from the drop down menu 3 Enter the label for the media as shown in Figure 4 18 It is best to use all capital letters with no spaces or punctuation marks Select Format 4 The system displays a pop...

Страница 103: ...1 Insert a formatted media into the drive 2 Enter enter Utility System Backup Restore NOTE If you are not logged in as GE Service or with administrator privileges the Operator Login window is displaye...

Страница 104: ...ator Login window is displayed Log on with administrator privileges 3 In the Restore list select Patient Archive Report Archive User Defined Configuration and Service 4 In the Media field select the B...

Страница 105: ...dia 2 To format the archive media enter Utility Connectivity Removable Media 3 Format the CD Verify the format if desired 4 Images will be moved from the hard drive by date Therefore the best way is t...

Страница 106: ...nformation on the first page of EZBack EZMove wizard then select Next If you want to backup all of the exams in the range even if the exam was previously backed up check this option If you uncheck thi...

Страница 107: ...ta and the storage Select Next to continue NOTE The calculation for the number of backup CD is only an estimate Allow for one additional CD when performing an EZBack EZMove NOTE This message Please in...

Страница 108: ...en complete press Next NOTE If you need to insert the next media a message appears providing you with the media label Label the media then insert the next media and press OK 10 When the backup is comp...

Страница 109: ...nguage settings Desired Language is displayed 4 Check assignment of Printer Keys The default function for Store and Print Keys are Store store image Print print Store and Print Keys can also be assign...

Страница 110: ...GE LOGIQ F SERIES DIRECTION 5446617 100 REVISION 10 BASIC SERVICE MANUAL 4 34 Section 4 5 Peripheral Checks This page was intentionally left blank...

Страница 111: ...apter 5 Components and Functions Theory Section 5 1 Overview 5 1 1 Purpose of Chapter 5 This chapter explains LOGIQ F SERIES system concepts component arrangement and subsystem function Table 5 1 Cont...

Страница 112: ...LOGIQ F SERIES DIRECTION 5446617 100 REVISION 10 BASIC SERVICE MANUAL 5 2 Chapter 5 Components and Functions Theory Section 5 2Block Diagram 5 2 1 System Diagram Figure 5 30 LOGIQ F Series System Diag...

Страница 113: ...GE LOGIQ F SERIES DIRECTION 5446617 100 REVISION 10 BASIC SERVICE MANUAL Chapter 5 Components and Functions Theory 5 3 5 2 2 Software Diagram Figure 5 31 LOGIQ F Series Software Diagram...

Страница 114: ...nd Functions Theory Section 5 3 Common Service Platform 5 3 1 Introduction The Service Platform contains a set of software modules that are common to all PC backend ultrasound and cardiology systems T...

Страница 115: ...6 Service Adjustments Section 6 1 Overview 6 1 1 Purpose of this chapter 6 This section describes how to test and adjust the scanner These tests are optional You may use them to check the system for...

Страница 116: ...LCD monitor s button located the right side of the LCD Monitor For LOGIQ C3 LC5 Premium Adjust the LCD monitor s button located the bottom of the LCD Monitor To adjust the light NOTE If the left side...

Страница 117: ...nd software that help maintain image quality and system operation Very basic host system and board level diagnostics are run whenever power is applied Some Service Tools may be run at the application...

Страница 118: ...ck to the manufacturer for analysis There are different options to acquire this data that would give different results 7 2 2 Collect Vital System Information The following information is necessary in...

Страница 119: ...Crash Log Power Supply Temperature NOTE Power Supply and Temperature logs are not currently being updated by the LOGIQ F Series This Alt D function is available at all times When Alt D is pressed a m...

Страница 120: ...mized settings 3 In the Destinations section record the service that is displayed 4 In the Physical Print Buttons section record the parameters related to the service 7 3 2 Setting the Print Key to Sc...

Страница 121: ...turing a Screen cont d 3 Click FREEZE to unfreeze the image to view the image screen and the snapshots displayed on the bottom 4 Highlight the snapshot to be stored to RD Removable Disk or CD R 5 Sele...

Страница 122: ...d in Setting the Print Key to Screen Capture on page 7 4 Refer to Figure 7 2 1 Click Utility on the keyboard 2 Select Connectivity from the Utilities Menu 3 Select the Buttons tab on the Connectivity...

Страница 123: ...ions 7 4 1 1 Disruptive Mode Allows you to enable or disable disruptive mode troubleshooting 7 4 1 2 System Shutdown Allows for system shutdown from the diagnostic menu Select to Restart System or Shu...

Страница 124: ...ts 7 4 2 2 Hard Drive Long 7 4 2 3 Hard Drive Short 7 4 2 4 Memory 7 4 2 5 Network Adapter 7 4 2 6 System Board 7 4 2 7 Video 7 4 3 PC Diagnostics Interactive Tests 7 4 3 1 AVI Playback 7 4 3 2 Keyboa...

Страница 125: ...h network 2 Enter Utility Connectivity TCP IP in IP settings window check Enable DHCP and select the proper network speed in Network Speed NOTE If user wants to setup static IP address uncheck Enable...

Страница 126: ...onfiguration 1 Connect system with network cont d 3 Select Save and a popup window displays Select OK to restart the system and activate the changes 4 After the system restarts the network icon at the...

Страница 127: ...1 1 Purpose of Chapter 8 This chapter describes replacement procedures for the following modules and subsystems Table 8 1 Contents in Chapter 8 Section Description Page Number 8 1 Overview 8 1 8 2 DI...

Страница 128: ...n is erased from the Harddisk Storage Device In case that any patient information is still residing on the LOGIQ F Series GE will contact the customer and request for urgent collection of that patient...

Страница 129: ...Down the System and disconnect the power cord 8 2 3 3 Removal Procedure 1 Pull out the air filter from the rear panel 2 Dust the filter with a vacuum cleaner and or wash it with a mild soapy solution...

Страница 130: ...ravel 8 3 0 3 Preparations Shut down the system 8 3 0 4 Procedure 1 Rotate the retainer counterclockwise until it can be removed from the keyboard refer to Figure 8 10 on page 8 4 2 Seperate the track...

Страница 131: ...eded 1 person 1minutes travel 8 4 0 4 Installation Procedure Attach the keyboard film to the keyboard 1 Use a cotton swab to clean around the keys NOTE Be careful and keep the keyboard film clean when...

Страница 132: ...8 6 Section 8 4 Attaching Keyboard Film Section 8 4 Attaching Keyboard Film cont d Fit the film on the keyboard well and then remove the remaining release paper 3 The keyboard film is attached success...

Страница 133: ...ers NOTE Please ensure power cable is connected during system upgrade 1 Insert the USB disk labeled System Application Software in the USB port of the system 2 Properly turn off the scanner by momenta...

Страница 134: ...Select one of the options for loading the system Select choice 1 to load the complete disk 7 Press any key to continue when below message display as shown below Figure 8 16 Selection for loading the...

Страница 135: ...e Software 8 9 Section 8 5 Loading Base Image Software cont d 8 System will be loaded as shown in the screen below 9 System updating finished refer to Figure 8 19 on page 8 9 10 Remove the USB stick t...

Страница 136: ...n off the system by momentarily pressing the Power On Off Switch Then select Shutdown from the System Exit Window NOTE If the system does not shutdown normally hold down the Power On Off Switch until...

Страница 137: ...he Power On Off key 7 Select one of the options and then press Enter to continue Select choice 1 to load the complete disk Refer to Figure 8 23 on page 8 11 To select 1 the complete disk will be loade...

Страница 138: ...re installation to complete The process is indicated on the screen 9 After the applying process completes the system begins to backup C Dsik WARNING While the software installation procedure is design...

Страница 139: ...IC SERVICE MANUAL Section 8 5 Loading Base Image Software 8 13 Section 8 5 Loading Base Image Software cont d 10 The upgrade process is complete the system will shut down automatically Refer to Figure...

Страница 140: ...GE LOGIQ F SERIES DIRECTION 5446617 100 REVISION 10 BASIC SERVICE MANUAL 8 14 Section 8 5 Loading Base Image Software This page was intentionally left blank...

Страница 141: ...tions Assy Assembly Ctrl Control FRU 1 Replacement part available in part hub FRU 2 Replacement part available from the manufacturer lead time involved Int Internal I O Input Output LCD Liquid Crystal...

Страница 142: ...China Class 1 1 AC Power Cord 5176773 2 Power cord India South Africa Class 1 1 AC Power Cord 5177195 2 Power cord Argentina Class 1 1 AC Power Cord 5176907 2 Power cord UK Class 1 1 AC Power Cord 517...

Страница 143: ...GE LOGIQ F SERIES DIRECTION 5446617 100 REVISION 10 BASIC SERVICE MANUAL Chapter 9 Renewal Parts 9 3 Section 9 4Operator Console Assy Figure 9 27 Operator Console Assy...

Страница 144: ...550 144 LOGIQ F Series User Guide Korean 1 N 6014 5453550 145 LOGIQ F Series User Guide Russian 1 N 6015 5453550 150 LOGIQ F Series User Guide Polish 1 N 6016 5453550 151 LOGIQ F Series User Guide Gre...

Страница 145: ...SION 10 BASIC SERVICE MANUAL Chapter 9 Renewal Parts 9 5 6033 5454731 141 LOGIQ F3 User Guide Chinese 1 N 6034 5457182 141 LOGIQ F5 F6 User Guide Chinese 1 N 6035 5444538 141 LOGIQ F8 User Guide Chine...

Страница 146: ...RS 5434194 Probe Center Frequency 6 5MHz 5499508 1 1 7001A 8C RS 5499508 Probe Center Frequency 6 5MHz 1 1 7002 E8C RS 5409293 Probe Center Frequency 6 5MHz 5499516 1 1 7002A E8C RS 5499516 Probe Cent...

Страница 147: ...and familiarize yourself with the procedures before performing a task Table 10 1 Contents in Chapter 10 Section Description Page Number 10 1 Overview 10 1 10 2 Why do Maintenance 10 2 10 3 Maintenanc...

Страница 148: ...ective action must be documented and maintained on the site Your GE service representative can help you with establishing performing and maintaining records for a quality assurance program Please cont...

Страница 149: ...eekly Inspect Cables and Connectors Clean Console Inspect Wheels Casters brakes and Swivel Locks Mobile Unit Check Daily Check Control Panel Movement Mobile Unit Check Daily Console Leakage Current Ch...

Страница 150: ...1 Screwdriver Blade Stubby 9 41322 Pickup Tool Claw type 9 41421 Screwdriver Blade Pocket clip 9 6757 6 pc Needle File Set 9 41594 Screwdriver Blade 1 8 in x 4 in 9 9487 Utility Knife 9 41581 Screwdri...

Страница 151: ...9 45381 Pliers Arc Joint 9 1 2 inch 9 44067 Socket 1 1 16 in for 1 2 in drive 9 45092 Pliers Linesman 8 1 2 inch 9 42679 Socket 10MM Hex for 1 2 in drive 2273333 9 42882 Punch Pin 3 32 inch 9 44262 E...

Страница 152: ...ound cord Anti Static Vacuum Cleaner 120V 230V Air Filter air intake Safety Analyzer The Safety Analyzer tool should be calibrated and compliant with AAMI ESI 1993 or IEC 60601 or AS NZS 3551 SVHS VCR...

Страница 153: ...hey may need to be soaked to remove excess gel Table 10 8 Physical Checks Step Item Description 1 Labeling Verify that all system labeling is present and in readable condition refer to the LOGIQ C3 C5...

Страница 154: ...the combination of lights that are illuminated Any problems found should be reported to the hospital immediately and the receptacle should not be used NOTE No outlet tester can detect the condition wh...

Страница 155: ...wire resistance 1 Turn the LOGIQ C3 C5 Premium unit OFF 2 Plug the unit into the meter and the meter into the tested AC wall outlet 3 Plug the black chassis cable into the meter s CHASSIS connector a...

Страница 156: ...unit ON and OFF with the power line polarity Normal and Reversed Record the highest reading of current When using the Microguard or a similar test instrument its power plug may be inserted into the w...

Страница 157: ...ghest leakage current measured 10 4 10 Isolated Patient Lead Sink Leakage Isolation Test reference the procedure in the IEC 60601 1 Select the ALL position on the lead selector Depress the rocker swit...

Страница 158: ...it GROUND OPEN GROUND CLOSED Patient Lead to Ground Leakage Current Test and Patient Lead to Lead Leakage Current Test 115V 10uA 10uA 220 240V 500uA 10uA Table 10 10 Maximum Allowance Limit for ECG Le...

Страница 159: ...ependent on its design Small variations in probe leakage currents are normal from probe to probe Other variations will result from differences in line voltage and test lead placement 10 4 11 3 Meter P...

Страница 160: ...lity Never switch the Polarity and the status of Neutral when the unit is powered ON Be sure to turn the unit power OFF before switching them using the POLARITY switch and or the NEUTRAL switch Otherw...

Страница 161: ...lies with the probe or the scanner NOTE Each probe will have some amount of leakage dependent on its design Small variations in probe leakage currents are normal from probe to probe Other variations w...

Страница 162: ...Scan Format Model Number Serial Number Probe 2 Frequency Scan Format Model Number Serial Number Probe 3 Frequency Scan Format Model Number Serial Number Probe 4 Frequency Scan Format Model Number Ser...

Страница 163: ...assis Source Leakage Current Probe Chassis Source Leakage Current Caster Chassis Source Leakage Current CRT Patient Lead Source Leakage Lead to Ground Patient Lead Source Leakage Lead to Lead Patient...

Страница 164: ...GE LOGIQ F SERIES DIRECTION 5446617 100 REVISION 10 BASIC SERVICE MANUAL 10 18 Section 10 5 When There s Too Much Leakage Current This page was intentionally left blank...

Страница 165: ...23 Control Panel 4 7 Monitor Display 4 7 4 10 Peripherals 4 33 Probes Connector Usage 4 24 G Gathering Trouble Data 7 2 General Cleaning 10 7 H Hazard Icons 1 7 Human Safety 1 8 L LOTO 1 12 M Mechani...

Страница 166: ...II Index T Touch Panel Functional Checks 4 7 Trouble Image with Logs 7 3 Troubleshooting Gathering Trouble Data 7 2 Screen Captures 7 4 Trouble Image with Logs 7 3 Vital System Information 7 2 U Ultra...

Страница 167: ...2013 2017 General Electric Company GE GEMedical Systems Ultrasound 9900 Innovation Drive Wawautosa Wisconsin 53226 USA...

Отзывы: