Section 7-3 - Screen Captures
Section 7-3
Screen Captures
There may be times when the customer or field engineer will want to capture a presentation on the
screen. This is accomplished by first saving the image(s) to the clipboard using a Print Key.
7-3-1
Check and Record the Store Key Function
Check the function of the Print Key in the event that the customer may have made some custom
settings.
1.) Enter
Utility
->
Connectivity
->
Buttons
.
2.) In the
Physical Print Buttons
field, select Print1, Print2 or Print3.
The Connectivity/Buttons Screen will be displayed like the one shown in
.
If Print key is not set to Whole Screen, as shown in
, proceed to step 5 to record the
customer’s customized settings.
3.) In the Destinations section, record the service that is displayed.
4.) In the
Physical Print Buttons
section, record the parameters related to the service.
7-3-2
Setting the Print Key to Screen Capture
If the Print Key is not set to screen capture:
1.) While on the Connect screen, with the Buttons tab displayed, go to the
Destinations
list.
2.) From the list select
Copy To Dataflow
. Press [>>] to add the selection to the
Printflow View
section.
3.) Ensure that the
Physical Print Buttons
section for capture Area is set to Whole Screen,
secondary Capture and No Image Compression.
4.) The Store Key should now be set up for whole screen capture, sending the screens to the
image buffer (clipboard).
7-3-3
Capturing a Screen
The following is a generic process to capture any screen from the scanner:
1.) Navigate to and display the image/screen to be captured.
2.) Press
. This will place a snapshot of the screen on the “clipboard” displayed at the bottom
of the scan image display.
Figure 7-2 Set print Key
Содержание LOGIQ F Series
Страница 1: ...GE LOGIQ F Series Basic Service Manual Part Number 5446617 100 Revision 10...
Страница 2: ......
Страница 3: ...GE LOGIQ F SERIES DIRECTION 5446617 100 REVISION 10 BASIC SERVICE MANUAL i Important Precautions...
Страница 4: ...GE LOGIQ F SERIES DIRECTION 5446617 100 REVISION 10 BASIC SERVICE MANUAL ii...
Страница 5: ...GE LOGIQ F SERIES DIRECTION 5446617 100 REVISION 10 BASIC SERVICE MANUAL iii...
Страница 6: ...GE LOGIQ F SERIES DIRECTION 5446617 100 REVISION 10 BASIC SERVICE MANUAL iv...
Страница 7: ...GE LOGIQ F SERIES DIRECTION 5446617 100 REVISION 10 BASIC SERVICE MANUAL v...
Страница 8: ...GE LOGIQ F SERIES DIRECTION 5446617 100 REVISION 10 BASIC SERVICE MANUAL vi...
Страница 9: ...GE LOGIQ F SERIES DIRECTION 5446617 100 REVISION 10 BASIC SERVICE MANUAL vii...
Страница 10: ...GE LOGIQ F SERIES DIRECTION 5446617 100 REVISION 10 BASIC SERVICE MANUAL viii...
Страница 11: ...GE LOGIQ F SERIES DIRECTION 5446617 100 REVISION 10 BASIC SERVICE MANUAL ix...
Страница 12: ...GE LOGIQ F SERIES DIRECTION 5446617 100 REVISION 10 BASIC SERVICE MANUAL x...
Страница 14: ...GE LOGIQ F SERIES DIRECTION 5446617 100 REVISION 10 BASIC SERVICE MANUAL xii...
Страница 167: ...2013 2017 General Electric Company GE GEMedical Systems Ultrasound 9900 Innovation Drive Wawautosa Wisconsin 53226 USA...