CHAPTER 4 FUNCTIONAL CHECK
4-1 Overview
In this chapter you will learn how to check all the main functions of LOGIQ A1 rapidly.
Please read the user manual of LOGIQ A1 to see how to check the rest functions
4-2 Procedures of Functional Check
4-2-1 Monitor Display after turning on
The test is performed with the linear probe when installing. Check these function again
when repairing the system.
The display of monitor is as follows (Refer Fig 4-1):
1. B mode image
2. Year, Month, Day, Hour, Minute
3. Total Gain (0—99)
4. Near Gain (-30—0)
5. Far Gain (0—30)
6. Focus points
7. Displaying mode and zooming
8. Scanning position
9. Measuring and Body marks
10. Measuring value
11. Probe kind
12. Frequency
13. Linear average
14. Frame average
15. Post-process
16. Depth control