GE MEDICAL SYSTEMS
LOGIQ A1 BASIC SERVICE MANUAL
REV 2
5176463-100
26
7. Body marks key 8. Frequency steps selection key 9. Image processing selecting key
10. Measuring key 11. Storage control key 12. Set key
13. Track ball 14. End key 15. Focus points selection key
16. Depth control key 17. White/black image reversing and scanning direction key
18. FREEZE key 19. Zooming key 20. B mode switching key
21. B/B mode switching key 22. B/M mode switching key 23. M mode switching key
24. Total gain selecting key 25. CAPSLOCK key 26. SHIFT key
27. Far gain selecting key 28. Near gain selecting key 29.
Clear key
30. Power indicator light
Illustration3-3 Control Panel
3-3 Electric performance
Power cord, tie-in, box, output socket, breaker and switches should be installed before
the mainframe of LOGIQ A1.
3-3-1 Power supply
All countries should install power cord according to their different requirements.
3-3-2 Electrical socket requirement
Separate electrical socket (220VAC) with 5A breaker is strongly recommended. This
kind of socket should adapt to the country’s criterion of power cord. It should be
approved by IEC or other related departments.
3-4 Storage and operating requirements
LOGIQ A1 is transported in separate package. It is about 10kg. Please find Table 3-2 to
see the temperature, atmospheric pressure and humidity range during transportation,
installation and operation.