Ingrid H.V. tank mounted on the Performix 160A X-Ray tube housing.
1.2.6 Description of the X-Ray head
The Innova System uses a Performix 160A X-Ray tube housing:
•
X-Ray tube
•
collimator
•
Ingrid HV tank
•
oil/water exchanger
•
contour filter
•
tube chiller
NOTE:
An external recirculating chiller is mandatory.
It is mandatory to place the Chiller in the Technical Room.
Illustration 2-10: Chiller Coolix 4100 (with auto-transformer)
1.2.7 Description of the Innova Imaging System
An Innova System is managed and controlled by a System including (does not list all
components):
•
C1 cabinet, inside which the main components are the RTAC, the DL, the Firewall, the HUB,
the KVM, the 4 KVA PDU and the Jedi generator,
•
C2 cabinet gantry and table control,
•
Fluoro UPS interface box near C1 cabinet (Fluoro UPS option),
•
VCIM console,
•
Revolution Digital Detector mounted on Gantry:
•
An external Digital Detector chiller is mandatory.
Innova
TM
IGS 520, Innova
TM
IGS 530, Innova
TM
IGS 540 Pre-Installation Manual
Direction 5499972-1-1EN, Revision 2
Chapter 2 Equipment Requirements