1.2.2.2 For the Fluoro UPS option
Table 5-5: Fluoro UPS option cables
Cable #
From
To
Type
Characteristics
Comments
1
Mains
PDB
3 phases
Diameter in conformity with
the Max Line Impedance ta‐
ble
Only with UL Fluoro UPS op‐
tion
Neutral
Diameter in conformity with
the Max Line Impedance ta‐
ble
Ground
AWG2/35 mm², and not
smaller than the neutral con‐
ductor
1
Mains
EMI Filter
3 phases
Diameter 35 mm² min, in
conformity with the Max Line
Impedance table
Only with CE Fluoro UPS op‐
tion
Neutral
Diameter 35 mm² min, in
conformity with the Max Line
Impedance table (35 mm²
min)
Ground
AWG2/35 mm² and not
smaller than the neutral con‐
ductor
8
EMI Filter
PDB
3 phases
Max length 3 m. Diameter in
conformity with the Max Line
Impedance table
Only with CE Fluoro UPS
Neutral
Max length 3 m. Diameter in
conformity with the Max Line
Impedance table
Ground
AWG2/35 mm², and not
smaller than the neutral con‐
ductor
9
PDB
Fluoro UPS
3 phases
10 mm² (CE)
Max diameter AWG3
AWG6 (UL)
Neutral
10 mm² (CE)
AWG6 (UL)
Ground
10 mm² (CE)
AWG6 (UL)
10
Fluoro UPS
PDB
3 phases
10 mm² (CE)
Max diameter AWG3
NOTE:
Cables 1 and 8 should be routed separately (CE Only) in order to avoid EMC
interference.
1.2.2.3 For the LDM option
Table 5-6: LDM option cables
Cable #
From
To
Type
Characteristics
Comments
11
Mains
LDM cabinet
Ground only
UL: AWG10 (5.26 mm²)
CE: AWG12 (2.5 mm²)
Innova
TM
IGS 520, Innova
TM
IGS 530, Innova
TM
IGS 540 Pre-Installation Manual
Direction 5499972-1-1EN, Revision 2
Chapter 5 Electrical Requirements
201