3.3.4 Injector Mounting Requirements
CAUTION
Table accessory rail load consideration:
The maximum load per table accessory rail is 40 kg (88 lbs) at 150 mm (0.49 ft)
(60 N.m or 44.25 ft/lbs). Therefore:
•
Only light extra load not exceeding 5 kg (11 lbs) at 100 mm (0.33 ft) (i.e IV
pole with its accessories, pressure head...) is authorized on the same table
accessory rail as the injector.
•
Never install injector and radiation protection on the same table accessory
rail.
•
Typical installation on the front table accessory rail is Smart handle or Smart
box, Table Side System Control (TSSC), Innova Central/Touchscreen, Table
panning device and cables support.
•
If needed an optional rail can be installed at table foot end of the Omega V
table for other options.
3.3.5 Prerequisites for ECG Acquisition Kit Installation
3.3.5.1 General Prerequisites for ECG Acquisition Kit Installation
Verify the following items to ensure easy mounting for the Hubican & Physio modules in all
situations:
1. Check enough space can be managed to properly install the Hubican module in Control
Room.
2. Check enough space can be managed to route/hide Hubican cable (VCIM-to-Hubican 12m
long cable) in Control Room.
3.3.5.2 Specific Prerequisites for Installation Configuration #1
Verify the following items only in the case of Installation Configuration #1 with ECG Device in
Control Room:
1. Check the environment of the ECG Device module located in Control Room provides
enough space to install the Physio module at a max distance of 1 meter.
2. Check enough space can be managed to properly route/hide Physio cable (Physio-to-
Hubican) in the Control Room
3. The Analog Output Box option is mandatory to provide on analog output connection to the
Physio module (If not present, it can be ordered through the following FRUs):
○
2018971-001 16CH ANALOG OUTPUT CPU INTERFACE OPTION
○
2007557-002 KIT ANALOG OUTPUT BOX W/CABLES
○
2010476-001 BOX CARDIOLAB/MACLAB ANALOG OUTPUT