The size of the air inlet has therefore to be dimensioned accordingly.
Contact your Local Distributor or one of the Service Centre, which will help you to find valuable
solutions.
The tables below indicate the heat dissipation at full load at PF = 0.8 lag. and charged battery,
up to 1000 m (3280 ft) altitude, for cooling air 25°C (77°F) to 30°C (86°F).
•
VFI (Voltage Frequency Independent) UPS system where the load is continuously supplied
by the inverter through the rectifier.
•
SEM (Super Eco Mode) permitting the maximum energy saving.
Table 4-5: Fluoro UPS CE
Losses
Cooling air flow
VFI
SEM
VFI
SEM
2.14 kW
0.64 kW
625 m
3
/h
190 m
3
/h
Table 4-6: Fluoro UPS UL
Losses
Cooling air flow
BTU / hr
kW
CFM
m
3
/h
6751
1.98
301
512
NOTICE
If installed in the technical room, the UPS may impact on it’s layout. It may also
be installed in a separate room. This depends on hospital constraints, local
regulations or EHS rules. Clearance, weight of UPS, airflow and cooling system
should be adapted for the UPS.
WARNING
ELECTRICAL CONTRACTOR IS RESPONSIBLE FOR PROVIDING AND
CONNECTING THE CABLES FROM THE PDB TO THE UPS AND
CONFIGURING THE PDB IN BYPASS MODE. GEHC IS RESPONSIBLE FOR
POWERING ON THE SYSTEM WITH THE UPS IN BYPASS MODE. GEDE IS
RESPONSIBLE FOR UPS COMMISSIONING.
CAUTION
The Fluoro UPS can be installed either in an Electrical Local with restricted
access or in the innova technical room as per local regulations.
NOTICE
Make sure that local regulations have been applied for the installation of the
Fluoro UPS (dedicated room/fire detection etc.)
Innova
TM
IGS 520, Innova
TM
IGS 530, Innova
TM
IGS 540 Pre-Installation Manual
Direction 5499972-1-1EN, Revision 2
192
3 Heat Output