Operation
Step
Select Quit from the File menu.
1
Select Stop the CCD cooling now and click OK.
2
TIP:
When Keep the CCD cooling after quit is selected, the cooling temperature of
the CCD will be maintained and the instrument can be used straight away the
next time the ImageQuant LAS 4000 Control Software is started. In this case, do
not turn off the ImageQuant LAS 4000. However, if the instrument will not be used
for a long period, it should be turned off.
Shut down the computer.
3
Switch the ImageQuant LAS 4000 off using the power switch.
4
4.15 Exposing large sample using wide view lens
A larger sample (of up to 250 x 250 mm) can be exposed by replacing a high-sensitivity lens with a
wide view lens. This section describes how to expose large samples.
Action
Step
Install the wide view lens. See
Section 3.2 Installing or exchanging the lens, on page 18
.
1
ImageQuant LAS 4000 User Manual 28-9607-42 AB
73
4 Operating the ImageQuant LAS 4000
4.14 Ending the session