4 Push the desired light source in along the guide at the bottom of the IDX.
5 Turn the lever at the front to lock the light source into position.
3.3.2 Changing the Epi lights
This section describes how to change a set of Epi lights. The RGB module Epi lights are also installed
as described here.
Note:
The tray should be at tray position 4 before installing Epi lights. If the tray is higher, then
it should be lowered to tray position 4 using the ImageQuant LAS 4000 Control Software
before attaching the lights, as described below.
Hook
Latch
Epi light base
1 If the tray is at Tray position 4, proceed to step 3. Otherwise, close the IDX door.
2 Click on Method/Tray position, set Tray position to 4, and click the OK button.
3 Open the IDX door. If no Epi light is currently in place inside the instrument, proceed to step 7.
ImageQuant LAS 4000 User Manual 28-9607-42 AB
29
3 Exchanging accessory parts
3.3 Changing the light sources
3.3.1 Changing the UV transilluminator or white light table