[EN] English – K0571 Issue 1
User Manual
Druck DPI611
4.4.1
Sequence to Upload and Download file
Table 4-1
Step
Procedure
1
Connect the USB cable (See
Section 1.11,
Accessories)
to the Druck DPI611 calibrator.
2
Connect the lead to the USB port on the computer
that has the calibrator manager installed.
3
Use 4 Sight to set-up the procedure and create a
work order for the device.
The procedure includes the parameters for the
calibration, the number of test points, the relationship
and pass/fail tolerance.
4
Use the Download button in calibrator manager to
Download the file to the Druck DPI611 calibrator. A
communications symbol will be displayed at the
bottom of the screen.
5
Select:
DASHBOARD >> DOCUMENTING >> RUN PROCEDURE
6
In the Results window, select the filename specified in
calibrator manager.
7
Enter the User ID and the DUT Serial Number.
8
Tap on the Start button
. The procedure sets up the
necessary Channel options e.g. mA and Volts.
9
Use the Take Reading button at each point specified
by the procedure. A prompt show for each point.
10
When all the readings are complete, tap on the Exit
Button
. Look at the results on the display
(As found/As Left).
11
To complete the process, use the calibrator manager
to Upload the file back into the 4 Sight database.