Druck DPI611
PERFORM CALIBRATION
Table 5-1
Before starting the calibration, check the time and date on the
instrument are correct.
To do a calibration on a measure or source function, use the
advanced menu option.
Select:
DASHBOARD >>
SETTINGS
>> ADVANCED
Enter the calibration PIN: 4321
Select the
button.
Select:
Then select a function and start the calibration:
1.
Select channel
2.
Select function
Function
Calibration equipment
(ppm = parts per million)
Current
Current (mA) calibrator.
Accuracy - Current measure/source,
See
Table 5-2
and
Table 5-3.
Voltage
Volts calibrator.
Accuracy - Voltage measure/source,
See
Table
5-5
or
Table 5-6.
Millivolts
mV calibrator.
Accuracy - Millivolts measure/source,
See Table
5-4
.
Pressure
(P1)
All Ranges total uncertainty of 0.025% reading or
better.
IDOS
UPM only. Refer to the user manual for the IDOS UPM.