22
Calibration
K0394 Issue 3 - [EN] English
V2 (two-point voltage
calibration)
Point 1 (P1)
- The DPI 104 shows the following displays:
1.
The calibration point to be used for
V2 - Point 1
.
After step
2
, the DPI 104 sets Vout to 0.1 V. Correct the value
(P1) to the value shown on the voltage calibrator.
5.
Repeat steps
3.
&
4.
for each digit. The value is ignored if it is
not in the permitted limits (Table 11, page 19).
Point 2 (P2)
- If P1 has a permitted value, the DPI 104 shows these
displays:
1.
The calibration point to be used for
V2 - Point 2
.
After step
2
, the DPI 104 sets Vout to 5.0 V. Correct the value
(P2) to the value shown on the voltage calibrator.
5.
Repeat steps
3.
&
4.
for each digit. The value is ignored if it is
not in the permitted limits (Table 11, page 19).
6.
If P2 has a permitted value, the DPI 104 uses the new P1/P2
values to adjust the output (
V
t
):
2.
3.
4.
8 seconds
0 to 9
2.
3.
4.
8 seconds
0 to 9
• Press this button to accept the
V2
calibration.
• The display shows “donE”.
• The instrument will restart.
• Press this button to reject the
V2
calibration and move to the next
menu option.